MUST2 SAMURAI_2018 S3-LEB-LPC SUPERNEMO MUGAST EXPAND SCALP GALATRON HiCARI VELO
  MUGAST, Page 42 of 132  Not logged in MUGAST
Status ExpNbr Date Author Type Category Target-Source Subjectdown
Fixede768s Fri Jul 12 16:35:10 2019 OttoRunRUNGeneral3-alphaRun303
*************************
Run : 303
*************************

Start at 2019-07-12 16:35:04.481793

3 alpha calibration run
+ MM2 masks
+ suppressed pedestals

Square detectors:
+ MG10: all X &Y strips masked;
+ MG9 removed

Stop at 2019-07-12 16:38:40.222071
Stop comment: type your message here
Fixede768s Fri Jul 12 16:42:22 2019 OttoRunRUNGeneral3-alphaRun303
JUNK
*************************
Run : 303
*************************

Start at 2019-07-12 16:41:50.518089

3 alpha calibration run
+ MM2 masks
+ suppressed pedestals (new pedestal for all trap detectors)

Square detectors:
+ MG10: not biased
+ MG9 removed

Stop at 2019-07-12 16:43:54.495178
Stop comment: type your message here
Fixede768s Thu Jul 11 16:54:34 2019 OttoRunRUNGeneral3-alphaRun302
*************************
Run : 302
*************************

Start at 2019-07-11 16:54:22.512184

3 alpha calibration run
+ without MM2 masks
+ suppressed pedestals

Square detectors:
+ MG10: all X strips masked; mate 14 & 16 masked
+ MG9: all X & Y strips masked

Stop at 2019-07-12 05:44:35.805449
Stop comment:
Fixede768s Mon Jul 15 14:38:27 2019 FrancoRUNGeneral3-alphaRun302
Attached the calibrated spectra for MUST2 and Mugast for run 302 (no pedestals).


OttoRun wrote:
*************************
Run : 302
*************************

Start at 2019-07-11 16:54:22.512184

3 alpha calibration run
+ without MM2 masks
+ suppressed pedestals

Square detectors:
+ MG10: all X strips masked; mate 14 & 16 masked
+ MG9: all X & Y strips masked

Stop at 2019-07-12 05:44:35.805449
Stop comment:
Attachment 1: run0302_must2_cal.png
run0302_must2_cal.png
Attachment 2: run0302_mugast_cal.png
run0302_mugast_cal.png
Fixede768s Fri Jul 19 23:10:18 2019 FrancoRUNGeneral3-alphaRun302
Added energy calibration for square detector (MG 10).


OttoRun wrote:
*************************
Run : 302
*************************

Start at 2019-07-11 16:54:22.512184

3 alpha calibration run
+ without MM2 masks
+ suppressed pedestals

Square detectors:
+ MG10: all X strips masked; mate 14 & 16 masked
+ MG9: all X & Y strips masked

Stop at 2019-07-12 05:44:35.805449
Stop comment:
Attachment 1: run302_MG_Ecal.png
run302_MG_Ecal.png
Fixede768s Fri Jul 19 23:12:36 2019 FrancoRUNGeneral3-alphaRun302
Added energy calibration for square detector (MG 10).


Franco wrote:
Attached the calibrated spectra for MUST2 and Mugast for run 302 (no pedestals).


OttoRun wrote:
*************************
Run : 302
*************************

Start at 2019-07-11 16:54:22.512184

3 alpha calibration run
+ without MM2 masks
+ suppressed pedestals

Square detectors:
+ MG10: all X strips masked; mate 14 & 16 masked
+ MG9: all X & Y strips masked

Stop at 2019-07-12 05:44:35.805449
Stop comment:
Attachment 1: run302_MG_Ecal.png
run302_MG_Ecal.png
Fixede768s Thu Jul 11 15:12:21 2019 OttoRunRUNGeneral3-alphaRun301
*************************
Run : 301
*************************

Start at 2019-07-11 15:07:27.670320

3 alpha calibration run
+ without MM2 masks
+ non suppressed pedestals

Square detectors:
+ MG10: all X strips masked; mate 14 & 16 masked
+ MG9: all X & Y strips masked

Stop at 2019-07-11 16:52:55.206383
Stop comment: stopping run to put pedestal suppression
Fixede768s Mon Jul 15 18:28:19 2019 FrancoRUNGeneral3-alphaRun301
Attached the calibrated spectra for MUST2 and Mugast for run 301.


OttoRun wrote:
*************************
Run : 301
*************************

Start at 2019-07-11 15:07:27.670320

3 alpha calibration run
+ without MM2 masks
+ non suppressed pedestals

Square detectors:
+ MG10: all X strips masked; mate 14 & 16 masked
+ MG9: all X & Y strips masked

Stop at 2019-07-11 16:52:55.206383
Stop comment: stopping run to put pedestal suppression
Attachment 1: run0301_must2_cal.png
run0301_must2_cal.png
Attachment 2: run0301_mugast_cal.png
run0301_mugast_cal.png
Fixede768s Tue Jul 16 09:51:56 2019 FrancoRUNGeneral3-alphaRun301
Attached the dat file with the position of the pedestals for each strip of MG3_X
that can be used to correct the pedestal problem of MG3_X in later runs (from run O306).



Franco wrote:
Attached the calibrated spectra for MUST2 and Mugast for run 301.


OttoRun wrote:
*************************
Run : 301
*************************

Start at 2019-07-11 15:07:27.670320

3 alpha calibration run
+ without MM2 masks
+ non suppressed pedestals

Square detectors:
+ MG10: all X strips masked; mate 14 & 16 masked
+ MG9: all X & Y strips masked

Stop at 2019-07-11 16:52:55.206383
Stop comment: stopping run to put pedestal suppression
Attachment 1: pedestals_MG3_STRX_E_RAW.dat
#Strip 	 Pedestal pos.
1	8192.3
2	8192.04
3	8192.01
4	8192.3
5	8192.07
6	8192.05
7	8192.53
8	8191.71
9	8191.33
10	8190.86
11	8190.64
12	8190.34
13	8189.44
14	8188.99
15	8437.76
16	8437.76
17	8192.91
18	8192.76
19	8192.5
20	8192.05
21	8192.05
22	8192.04
23	8191.66
24	8191.9
25	8191.88
26	8191.93
27	8191.77
28	8191.15
29	8191.19
30	8191.39
31	8191.63
32	8192.03
33	8192.28
34	8192.46
35	8191.93
36	8191.42
37	8192.15
38	8192.55
39	8192.74
40	8192.58
41	8192.57
42	8193.06
43	8192.91
44	8192.55
45	8192.46
46	8192.66
47	8192.61
48	8192.83
49	8192.98
50	8192.61
51	8192.45
52	8192.87
53	8192.83
54	8192.57
55	8192.57
56	8192.46
57	8192.59
58	8192.48
59	8192.53
60	8192.54
61	8192.65
62	8192.68
63	8192.57
64	8192.7
65	8191.82
66	8190.82
67	8191.16
68	8191.34
69	8191.22
70	8191.68
71	8191.89
72	8191.74
73	8191.69
74	8192.04
75	8191.72
76	8191.75
77	8191.83
78	8192
79	8191.93
80	8191.85
81	8192.37
82	8192.54
83	8192.52
84	8192.29
85	8192.19
86	8191.81
87	8192.06
88	8192.51
89	8192.55
90	8192.27
91	8192.24
92	8192.35
93	8192.07
94	8192.51
95	8191.28
96	8190.31
97	8191.3
98	8191.94
99	8192.28
100	8192.65
101	8192.4
102	8191.86
103	8191.55
104	8191.57
105	8191.67
106	8191.76
107	8191.97
108	8191.58
109	8191.6
110	8191.96
111	8191.77
112	8191.89
113	8192.44
114	8192.58
115	8192.51
116	8192.37
117	8192.06
118	8191.6
119	8191.88
120	8191.7
121	8191.5
122	8191.73
123	8191.6
124	8191.4
125	8191.67
126	8191.94
127	8191.52
128	8191.83
Fixede768s Tue Jul 16 09:17:48 2019 OttoRunRUNGeneralN/ARun30
*************************
Run : 30
*************************

Start at 2019-07-16 00:54:27.174959

fsc / calibT run

internal pulser frequency now set to 300 Hz (was about 1 kHz before)

GMT: MUGAST (square detector not biased) + MUST2

Time calibrator:
range -> 640 ns
freq. -> 20 ns

fsc modified to wait 60s instead of 10s
Fixede768s Tue Jul 16 09:18:07 2019 OttoRunRUNGeneralN/ARun29
*************************
Run : 29
*************************

Start at 2019-07-16 00:35:09.645821

fsc / calibT run

internal pulser frequency now set to 300 Hz (was about 1 kHz before)

GMT: MUGAST (square detector not biased) + MUST2

Time calibrator:
range -> 640 ns
freq. -> 20 ns

fsc modified to wait 60s instead of 10s
Fixede786s Sun Jun 23 18:09:17 2019 OttoRunRUNGeneralCryo-emptyRun28
*************************
Run : 28
*************************

Start at 2019-06-23 18:09:16.019140

VAMOS tunning Continued with 2.5 10*4
After Fixing the IC preamp issue
Brho0=1.05 Tm

DC pressure 6 mbar
IC pressure 120 mbar

CATS2 inserted


TRIGGER
VAMOS %1000
MUGAST
MUS2

Stop at 2019-06-23 18:15:20.785307
Stop comment:
Fixede768s Tue Jul 16 09:17:49 2019 OttoRunRUNGeneralN/ARun28
*************************
Run : 28
*************************

Start at 2019-07-16 00:15:50.617493

fsc / calibT run

internal pulser frequency now set to 300 Hz (was about 1 kHz before)

GMT: MUGAST (square detector not biased) + MUST2

Time calibrator:
range -> 640 ns
freq. -> 20 ns

fsc modified to wait 60s instead of 10s

Stop at 2019-07-16 09:17:51.288072
Stop comment:

Stop at 2019-07-16 09:17:51.824283
Stop comment:

Stop at 2019-07-16 09:17:52.302740
Stop comment:

Stop at 2019-07-16 09:17:52.781598
Stop comment:
Fixede768s Tue Jul 16 09:18:09 2019 OttoRunRUNGeneralN/ARun27
*************************
Run : 27
*************************

Start at 2019-07-15 23:56:24.751783

fsc / calibT run

internal pulser frequency now set to 300 Hz (was about 1 kHz before)

GMT: MUGAST (square detector not biased) + MUST2

Time calibrator:
range -> 640 ns
freq. -> 20 ns

fsc modified to wait 60s instead of 10s
Fixedcom2019 Sun Apr 7 21:45:35 2019 OttoRunRUNGeneralCD2Run262
*************************
Run : 262
*************************

Start at 2019-04-07 21:45:06.409673

Run Conditions:
  • Operator: Iulian Marlene
  • Trigger: MUST2 + MG BWd + VAMOS % 100
  • Target/Source: CD2 1mg
  • Beam: no
  • Intensity:

Comment: Target holder removed
background check


Stop at 2019-04-08 07:38:35.058779
Stop comment:
end run
Alfa background is continuous in energy particularly in the backward detectors


Attachment 1: bilan_run262_bck_mugast4.png
bilan_run262_bck_mugast4.png
Attachment 2: bilan_run262_bck_mugast3.png
bilan_run262_bck_mugast3.png
Attachment 3: bilan_run262_bck_mugast2.png
bilan_run262_bck_mugast2.png
Attachment 4: bilan_run262_bck_mugast1.png
bilan_run262_bck_mugast1.png
Fixedcom2019 Sun Apr 7 21:43:29 2019 OttoRunRUNGeneralCD2Run261
*************************
Run : 261
*************************

Start at 2019-04-07 21:43:00.572161

Run Conditions:
  • Operator: Iulian Marlene
  • Trigger: MUST2 + MG BWd + VAMOS % 100
  • Target/Source: CD2 1mg
  • Beam: no
  • Intensity:

Comment: Target holder in position


junk

Stop at 2019-04-07 21:44:59.581974
Stop comment: junk
Fixede744s Wed Apr 10 10:36:48 2019 OttoRunRUNGeneralN/ARun26
*************************
Run : 26
*************************

Start at 2019-04-10 10:36:46.761926

Run Conditions:
  • Operator: A. Lemasson
  • Trigger: Vamos Pb with DC 63
  • Target/Source: none
  • Beam: VAMOS Closed
  • Intensity:

Comment:
Test ACQ


Stop at 2019-04-10 10:46:01.017415
Stop comment:
Fixede786s Sun Jun 23 17:44:22 2019 OttoRunRUNGeneralCryo-emptyRun26
*************************
Run : 26
*************************

Start at 2019-06-23 17:44:18.480300

VAMOS tunning Continued with 2.5 10*4
After Fixing the IC preamp issue
Brho0=1.05 Tm

DC pressure 6 mbar
IC pressure 120 mbar

CATS2 inserted


TRIGGER
VAMOS %1000
MUGAST
MUS2

Stop at 2019-06-23 18:03:57.967230
Stop comment:

Stop at 2019-06-23 18:08:11.630931
Stop comment:
Fixede768s Tue Jul 16 09:32:10 2019 OttoRunRUNGeneralN/ARun26
*************************
Run : 26
*************************

Start at 2019-07-15 23:37:09.715822

fsc / calibT run

internal pulser frequency now set to 300 Hz (was about 1 kHz before)

GMT: MUGAST (square detector not biased) + MUST2

Time calibrator:
range -> 640 ns
freq. -> 20 ns

fsc modified to wait 60s instead of 10s
Fixedcom2019 Sat Apr 6 18:32:26 2019 OttoRunRUNGeneralCD2Run259
*************************
Run : 259
*************************

Start at 2019-04-06 18:31:46.303930

Run Conditions:
  • Operator:
  • Trigger: MUST2 + MG BWd + VAMOS % 100
  • Target/Source: CD2 1mg
  • Beam: no
  • Intensity:

Comment: Target holder in position

ELOG Home Page