MUST2 SAMURAI_2018 S3-LEB-LPC SUPERNEMO MUGAST EXPAND SCALP GALATRON HiCARI VELO
  MUGAST, Page 3 of 132  Not logged in MUGAST
Fixed, e793s, Mon Mar 8 18:07:25 2021, mugast, GRIT, General, N/A, Run 633 - TACs calibration (part 3) 
TAC calibration.

Mugast-CATS2 (range 1 us)
Mugast-CATS2b (range 2 us)
AGATA-CATS2 (for this one better to use run 634)

Time calibrator

2.56 us range
40 ns steps
Fixed, e793s, Mon Mar 8 18:15:07 2021, mugast, GRIT, General, N/A, Run 634 - TACs calibration (part 4) 
TAC calibration.

AGATA-CATS2 (range 200 ns)

Time calibrator

2.56 us range
10 ns steps
Fixed, e793s, Mon Mar 8 19:01:26 2021, Franco, GRIT, General, 3-alpha, Energy and time calibrations before experiment 8x
Energy resolutions:

MG1: 45 keV
MG2: 45 keV
MG3: 45 keV
MG4: 50 keV
MG5: 50 keV
MG7: 50 keV
Total Mugast: 50 keV

MM1: 50 keV
MM2: 50 keV
MM3: 45 keV
MM4: 50 keV
Total MUST2: 50 keV

MM5 (90 deg): 65 keV
Pinned, e793s, Tue Mar 9 09:35:18 2021, Adrien, OTHER, General, N/A, E793s proposal Proposal_K48_by_dp_MUGAST_FINAL.pdf
Attached is the proposal of the experiment.
Fixed, e793s, Tue Mar 9 09:43:51 2021, OttoRun, RUN, General, 3-alpha, Run200 
*************************
Run : 200
*************************

Start at 2021-03-08 17:43:53.831102

run 3 alpha

Stop at 2021-03-09 09:43:55.629095
Stop comment: type your message here

Stop at 2021-03-09 09:43:56.774034
Stop comment: type your message here

Stop at 2021-03-09 09:43:57.137511
Stop comment: type your message here

Stop at 2021-03-09 09:43:57.480913
Stop comment: type your message here
Fixed, e793s, Tue Mar 9 10:22:16 2021, mugast, DAQ, Hardware, N/A, U2M scaler configuration 
U2M

1-FAG
2-OR MM-MG
3-OR AGATA
4-TRIG VAMOS
5-CATS CLEAN
6-DT && OR MM-MG
7-100 Hz
8-100 Hz & DT
9-MUVI1
10-MUVI2
11-MUVI3
12-MUVI4
15-CATS2 direct
16- FINGER (unused)
Fixed, e793s, Tue Mar 9 10:47:12 2021, mugast, DAQ, Hardware, N/A, GMT config 
GMT

1-MUVI1
2-MUVI2
3-MUVI3
4-
5-CATS2clean % (/10**4 for the moment)
6-VAMOS% (/10 for the moment)
7-MUST2% (/10 for the moment)
8-
9-
10-VAMOS
11-OR AGATA
Fixed, e793s, Tue Mar 9 11:47:14 2021, Franco, RUN, General, N/A, TACS calibration TACS_calib.pngTACS_calib_fits.png
Coefficients uploaded in MUGAST/e793s/calibTACS_e793s.cal (set offsets to 0 and kept only the slope)
Fixed, e793s, Wed Mar 10 13:03:00 2021, OttoRun, RUN, General, 3-alpha, Run41 
*************************
Run : 41
*************************

Start at 2021-03-10 12:58:41.805693

run 3 alpha

Stop at 2021-03-10 13:06:49.699879
Stop comment:
Fixed, e793s, Wed Mar 10 13:09:12 2021, OttoRun, RUN, General, 3-alpha, Run42 
*************************
Run : 42
*************************

Start at 2021-03-10 13:09:04.739262

run 3 alpha

Stop at 2021-03-10 13:30:31.317348
Stop comment: changing trigger
Pinned, e793s, Wed Mar 10 13:10:14 2021, Adrien, BEAM, General, CD2, Kinematical line from Ti contamination KineTiK.pdfKine.cxx
In black the Kinematical line for states known to be populated strongly in 47Ti(d,p).
In blue range of the states of interrest.

The contamination is in a different region and should not cause any issue.
Fixed, e793s, Wed Mar 10 13:33:31 2021, OttoRun, RUN, General, N/A, Run43 
*************************
Run : 43
*************************

Start at 2021-03-10 13:33:01.346673

Agata or as trigger for testing before beam

Stop at 2021-03-10 13:55:37.322216
Stop comment:
Fixed, e793s, Wed Mar 10 13:57:54 2021, OttoRun, RUN, General, N/A, Run45 
*************************
Run : 45
*************************

Start at 2021-03-10 13:57:50.701600

Agata or as trigger for testing before beam

Stop at 2021-03-10 14:01:53.157232
Stop comment:
Fixed, e793s, Wed Mar 10 14:43:14 2021, OttoRun, RUN, General, N/A, Run46 
*************************
Run : 46
*************************

Start at 2021-03-10 14:42:50.365714

Test run for Professor Clement. Beam in the room.

Stop at 2021-03-10 16:02:30.981195
Stop comment:
Fixed, e793s, Wed Mar 10 15:15:44 2021, Charlie, OTHER, General, , 48Ti Kinematics (additional) PDBarnes_Table_1.png48Ti_and_48K_Kinematics.pngWilton_mock-up.png
Supplementary to Adrien's kinematic calculations of 47Ti(d,p)48Ti, we have found a paper suggesting that by far the most strongly populated state in 48Ti is at 5.65 MeV. There is also a doublet of strong states at around 4.4 MeV, and two more strongly populated states at 6.06 and 6.33 MeV.

See attachment for the full table from P. D. Barnes (1965), Phys.Rev.138.B597.

Some quick CatKin models also show that a (theoretical) state at 7 MeV in 48Ti would almost perfectly overlap with the ground state of 48K. As such, when looking at the lab energy with the g.s. of 48K normalised to 0, we could expect to see these strongly populated 48Ti states at -0.67, -0.94, -1.35 and -2.6 MeV (for 6.3, 6.06, 5.65 and 4.4 MeV, respectively).

See the attached plot for kinematics of 47K(d,p)48K @ 0 MeV and 4 MeV, as well as 47Ti(d,p)48Ti @ 0 MeV, 5.65 MeV and the theoretical 7 MeV. Additionally, see a mock-up of the expected spectra from Wilton.
Fixed, e793s, Wed Mar 10 16:07:43 2021, OttoRun, RUN, General, N/A, Run47 
*************************
Run : 47
*************************

Start at 2021-03-10 16:07:22.660368

Beam in the room; restarting daq after recompiling the action

Stop at 2021-03-10 16:09:55.182936
Stop comment: stopping run to remove vamos% from trigger
Fixed, e793s, Wed Mar 10 16:12:40 2021, OttoRun, RUN, General, Empty, Run48 
*************************
Run : 48
*************************

Start at 2021-03-10 16:11:56.052849

Beam in the room; restarting daq after changing trigger.

Trigger are MUGAST/MM/CATS%

Stop at 2021-03-10 16:30:03.651340
Stop comment: stopping to remove cats% from trigger so we can set the timing
Fixed, e793s, Wed Mar 10 16:31:50 2021, OttoRun, RUN, General, CD2, Run49 
*************************
Run : 49
*************************

Start at 2021-03-10 16:31:31.923710

Beam in the room; restarting daq after changing trigger.
target: CD2 0.5mg
Trigger are MUGAST/MM

Stop at 2021-03-10 16:53:12.237279
Stop comment: stopping to change GMT windows
Fixed, e793s, Wed Mar 10 17:00:06 2021, OttoRun, RUN, General, CD2, Run50 
*************************
Run : 50
*************************

Start at 2021-03-10 16:59:44.368649

Beam in the room; restarting daq after changing trigger.
target: CD2 0.5mg
Trigger are MUGAST/MM
changing width of the FAG to 1us

Stop at 2021-03-10 17:12:21.612357
Stop comment: stopping to get in and change the cats delay
Fixed, e793s, Wed Mar 10 17:25:22 2021, OttoRun, RUN, General, CD2, Run51 
*************************
Run : 51
*************************

Start at 2021-03-10 17:24:45.316220

Beam in the room; restarting daq after changing trigger.
target: CD2 0.5mg
Trigger are MUGAST/MM
changed width of the FAG to 400ns
Removed extra 100ns delay from cats validation

Stop at 2021-03-10 18:14:26.996869
Stop comment: stop for VAMOS data analysis. Coherence check

Stop at 2021-03-10 20:46:36.348405
Stop comment: stopped the run to add cats% in the trigger

ELOG Home Page