Status |
ExpNbr |
Date |
Author |
Type |
Category |
Target-Source |
Subject |
Fixed | e786s |
Sat Jun 29 05:56:40 2019 |
OttoRun | RUN | General | Cryo-3He | Run77 |
OttoRun wrote: |
OttoRun wrote: |
OttoRun wrote: | *************************
Run : 77
*************************
Start at 2019-06-28 18:34:22.873862
target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted
AGATA position +51-25 mm
TRIGGER :
- CMG
- M2
- VAMOS%1000
- CATS%1000
all X on MUVI disabled from trigger
all threshold of MUVI1 at 1 MeV to avoid as much as possible noise
beam back. Q pole issues still |
Beam is back, again, at 1.53, weird behavior of CATS |
|
After beam is back we found 350Hz on vamos focal plane, with 20kHz current. After Brho scanning we found a Brho distribution going from 1Tm to 0.4Tm. Observation of gamma rays indicates no accumulation of radioactivity in the chamber with respect to previous runs. We conclude that we may have ice layers up to 500um in the target but very in-homogeneous since we also observe 1Tm Brho. To verify maybe increase target pressure to modify the structure of the ice. CATS is tripping continuously. No VACUUM changes observed. |
Fixed | e768s |
Thu Jul 18 00:16:56 2019 |
OttoRun | RUN | General | N/A | Run77 |
*************************
Run : 77
*************************
Start at 2019-07-18 00:11:48.462091
Time calibrator starting from MG3
strip by strip
Range 640 ns
Period 160 ns |
Fixed | e744s |
Thu Apr 11 08:01:12 2019 |
OttoRun | RUN | General | CH2 | Run76 |
*************************
Run : 76
*************************
Start at 2019-04-11 07:59:50.438662
Run Conditions:
- Operator: Nikola, Marion
- Trigger: Mugast & Vamos % 1000 & Cats2 % 40000
- Target/Source: CH2 105 um (Position 2)
- Beam: 14N
- Intensity: ~ 2e+5
- Brho = 0.437 Tm
- CATS 2 IN
Comment: Beam stopped (Balise).
Stop at 2019-04-11 08:33:56.045621
Stop comment: |
Fixed | e786s |
Fri Jun 28 15:25:12 2019 |
OttoRun | RUN | General | Cryo-3He | Run76 |
*************************
Run : 76
*************************
Start at 2019-06-28 15:24:51.867946
target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted
AGATA position +51-25 mm
TRIGGER :
- CMG
- M2
- VAMOS%1000
- CATS%1000
all X on MUVI disabled from trigger
all threshold of MUVI1 at 1 MeV to avoid as much as possible noise
beam back. Q pole issues still
Stop at 2019-06-28 16:55:27.258794
Stop comment: |
Fixed | e768s |
Thu Jul 18 00:16:52 2019 |
OttoRun | RUN | General | N/A | Run76 |
*************************
Run : 76
*************************
Start at 2019-07-17 23:09:25.888318
Time calibrator starting from MG3
strip by strip
Range 640 ns
Period 160 ns |
Fixed | e744s |
Thu Apr 11 06:41:14 2019 |
OttoRun | RUN | General | CH2 | Run75 |
========================================
*************************
Run : 75
*************************
Start at 2019-04-11 06:40:54.479084
Run Conditions:
- Operator: A. Lemasson
- Trigger: Mugast & Vamos % 1000 & Cats2 % 40000
- Target/Source: CH2 105 um (Position 2)
- Beam: 14N
- Intensity: ~ 2e+5
- Brho = 0.437 Tm
- CATS 2 IN
Comment:
Stop at 2019-04-11 07:29:11.245553
Stop comment: |
Fixed | e744s |
Thu Apr 11 07:17:33 2019 |
OttoRun | RUN | General | CH2 | Run75 |
OttoRun wrote: | *************************
Run : 75
*************************
Start at 2019-04-11 06:40:54.479084
Run Conditions:
- Operator: A. Lemasson
- Trigger: Mugast & Vamos % 1000 & Cats2 % 40000
- Target/Source: CH2 105 um (Position 2)
- Beam: 14N
- Intensity: ~ 2e+5
- Brho = 0.437 Tm
- CATS 2 IN
Comment: |
|
Attachment 1: run75_1.png
|
|
Attachment 2: run75_2.png
|
|
Attachment 3: run75_3.png
|
|
Attachment 4: run75_4.png
|
|
Attachment 5: run75_5.png
|
|
Attachment 6: run75_6.png
|
|
Attachment 7: run75_7.png
|
|
Attachment 8: run75_8.png
|
|
Attachment 9: run75_9.png
|
|
Attachment 10: run75_10.png
|
|
Attachment 11: run75_11.png
|
|
Attachment 12: run75_scaler.png
|
|
Fixed | e786s |
Fri Jun 28 10:00:25 2019 |
OttoRun | RUN | General | Cryo-3He | Run75 |
*************************
Run : 75
*************************
Start at 2019-06-28 09:59:59.505010
target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted
AGATA position +51-25 mm
TRIGGER :
- CMG
- M2
- VAMOS%1000
- CATS%1000
all X on MUVI disabled from trigger
all threshold of MUVI1 at 1 MeV to avoid as much as possible noise
CATS threshold raised
Stop at 2019-06-28 14:33:09.843668
Stop comment: |
Fixed | e768s |
Thu Jul 18 00:16:58 2019 |
OttoRun | RUN | General | N/A | Run75 |
*************************
Run : 75
*************************
Start at 2019-07-17 22:06:59.976407
Time calibrator starting from MG3
strip by strip
Range 640 ns
Period 160 ns
Stop at 2019-07-18 00:16:59.664044
Stop comment:
Stop at 2019-07-18 00:17:01.189328
Stop comment:
Stop at 2019-07-18 10:44:37.670485
Stop comment: |
Fixed | e744s |
Thu Apr 11 04:39:48 2019 |
OttoRun | RUN | General | CH2 | Run74 |
*************************
Run : 74
*************************
Start at 2019-04-11 04:37:42.487024
Run Conditions:
- Operator: AIulian Marlene Diego
- Trigger: Mugast & Vamos % 1000 & Cats2 % 40000
- Target/Source: CH2 105 um (Position 2)
- Beam: 14N
- Intensity: ~ 2e+5
- Brho = 0.437 Tm
- CATS 2 IN
First good data run. 2.7e+5 counts Cats2
Comment:
Stop at 2019-04-11 06:23:39.836186
Stop comment: good run |
Fixed | e786s |
Thu Jun 27 23:28:30 2019 |
OttoRun | RUN | General | Cryo-3He | Run74 |
*************************
Run : 74
*************************
Start at 2019-06-27 23:25:00.455575
target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted
AGATA position +51-25 mm
TRIGGER :
- CMG
- M2
- VAMOS%1000
- CATS%1000
all X on MUVI disabled from trigger
all threshold of MUVI1 at 1 MeV to avoid as much as possible noise
Stop at 2019-06-28 09:31:40.775271
Stop comment: |
Fixed | e768s |
Thu Jul 18 00:16:53 2019 |
OttoRun | RUN | General | N/A | Run74 |
*************************
Run : 74
*************************
Start at 2019-07-17 21:03:59.134373
Time calibrator starting from MG3
strip by strip
Range 640 ns
Period 160 ns
Stop at 2019-07-18 00:16:55.063278
Stop comment: |
Fixed | e744s |
Thu Apr 11 03:40:44 2019 |
OttoRun | RUN | General | CH2 | Run73 |
*************************
Run : 73
*************************
Start at 2019-04-11 03:39:15.971913
Run Conditions:
- Operator: A. Lemasson
- Trigger: VAMOS
- Target/Source: CH2 105 um (Position 2)
- Beam: 14N
- Intensity: ~ 2e+4
- Brho = 0.437 Tm
- CATS 2 IN
Comment:
Stop at 2019-04-11 04:25:03.931527
Stop comment: |
Fixed | e786s |
Thu Jun 27 20:26:27 2019 |
OttoRun | RUN | General | Cryo-3He | Run73 |
*************************
Run : 73
*************************
Start at 2019-06-27 20:26:17.916314
target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted
AGATA position +51-25 mm
TRIGGER :
- CMG
- M2
- VAMOS%1000
- CATS%1000
all X on MUVI disabled from trigger
all threshold of MUVI1 at 1 MeV to avoid as much as possible noise
There is no run 72
Stop at 2019-06-27 23:04:45.884452
Stop comment: AGATA DAQ is rejectting all events |
Fixed | e768s |
Wed Jul 17 20:01:57 2019 |
OttoRun | RUN | General | N/A | Run73 |
*************************
Run : 73
*************************
Start at 2019-07-17 20:00:28.897236
Time calibrator starting from MG3
strip by strip
Range 640 ns
Period 160 ns
Stop at 2019-07-18 00:16:51.295899
Stop comment: |
Fixed | e744s |
Thu Apr 11 03:21:14 2019 |
OttoRun | RUN | General | CH2 | Run72 |
*************************
Run : 72
*************************
Start at 2019-04-11 03:16:52.893246
Run Conditions:
- Operator: A. Lemasson
- Trigger: VAMOS
- Target/Source: CH2 105 um (Position 2)
- Beam: 14N
- Intensity: ~ 2e+4
- Brho = 0.486 Tm
- CATS 2 IN
Comment:
DAQ Crash |
Fixed | e768s |
Wed Jul 17 17:18:37 2019 |
OttoRun | RUN | General | LiF | Run72 |
*************************
Run : 72
*************************
Start at 2019-07-17 17:11:53.463967
15N 1e8 pps
LiF target
Stop at 2019-07-17 19:39:32.941104
Stop comment: |
Fixed | e744s |
Thu Apr 11 03:09:14 2019 |
OttoRun | RUN | General | CH2 | Run71 |
*************************
Run : 71
*************************
Start at 2019-04-11 03:08:43.011636
Run Conditions:
- Operator: A. Lemasson
- Trigger: VAMOS
- Target/Source: CH2 105 um (Position 2)
- Beam: 14N
- Intensity: ~ 2e+4
- Brho = 0.484 Tm
- CATS 2 IN
Comment:
Changed Brho to 0.484 to move the beam oit
Stop on DAQ Crash
(03:11:50 - 11/04/2019) FATAL - log from NARVAL.NUMEXO2_61 - Bloc_Catcher.T_Communication : requested size too high available in actor 1024 requested in header 5376 do not use this actor
(03:11:50 - 11/04/2019) FATAL - log from NARVAL.NUMEXO2_61 - NUMEXO2_61 put sub system NARVAL in error with message :Bloc_Catcher.T_Communication : actor buffer size too small
(03:11:50 - 11/04/2019) ERROR - log from NARVAL.NUMEXO2_61 - Bloc_Catcher.T_Communication : BlC : exception,STORAGE_ERROR,s-intman.adb:136 explicit raise
(03:11:50 - 11/04/2019) ERROR - log from NARVAL.NUMEXO2_61 - Bloc_Catcher.T_Communication : BPR2 : exception,STORAGE_ERROR,s-intman.adb:136 explicit raise
(03:11:50 - 11/04/2019) ERROR - log from NARVAL.NUMEXO2_61 - Bloc_Catcher.T_Communication : BPR ,STORAGE_ERROR,s-intman.adb:136 explicit raise
(03:11:50 - 11/04/2019) ERROR - log from NARVAL.NUMEXO2_61 - Bloc_Catcher.T_Communication : exception ,STORAGE_ERROR,s-intman.adb:136 explicit raise
Stop at 2019-04-11 03:20:00.545286
Stop comment: |
Fixed | e786s |
Thu Jun 27 18:48:51 2019 |
OttoRun | RUN | General | Cryo-3He | Run71 |
*************************
Run : 71
*************************
Start at 2019-06-27 18:47:56.171341
target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted
AGATA position +51-25 mm
TRIGGER :
- MG
- square (not biased)
- M2
all X on MUVI disabled from trigger
all threshold of MUVI1 at 1 MeV to avoid as much as possible noise
Stop at 2019-06-27 20:23:56.630240
Stop comment: |
Fixed | e768s |
Wed Jul 17 13:55:30 2019 |
OttoRun | RUN | General | LiF | Run71 |
*************************
Run : 71
*************************
Start at 2019-07-17 13:54:44.082989
15N beam -- data tacking
Target #1 1.25 mg/cm2
Brho = 0.640 Tm
plastic off
diamond on
mask: +7 mm
no beam at the beginning; we are requesting it.
restart daq qfter chqngin MUST2 HV mainframe.
Stop at 2019-07-17 14:49:13.546336
Stop comment: beam is taken back |