Status |
ExpNbr |
Date |
Author |
Type |
Category |
Target-Source |
Subject |
Fixed | e744s |
Thu Apr 11 19:49:29 2019 |
OttoRun | RUN | General | Empty | Run82 |
*************************
Run : 82
*************************
Start at 2019-04-11 19:49:24.212733
Run Conditions:
- Operator: Adrien
- Trigger: VAMOS %1
- Target/Source: no target
- Beam: 14O
- Intensity: ~ 5e+3
- Brho = 0.69186 Tm
- !CATS/VAMOS (CATS Efficiency)! : XXXX
Comment:
Aligning beam, daq on for online monitoring
Stop at 2019-04-11 19:54:51.148174
Stop comment: 5 PI emittance for 14O beam |
Fixed | e744s |
Thu Apr 11 19:09:35 2019 |
OttoRun | RUN | General | Empty | Run81 |
*************************
Run : 81
*************************
Start at 2019-04-11 19:09:20.734153
Run Conditions:
- Operator: Adrien
- Trigger: VAMOS %1
- Target/Source: no target
- Beam: 14O
- Intensity: ~ 5e+3
- Brho = 0.69186 Tm
- !CATS/VAMOS (CATS Efficiency)! : XXXX
Comment:
Aligning beam, daq on for online monitoring
Stop at 2019-04-11 19:49:20.042737
Stop comment: |
Fixed | e786s |
Sat Jun 29 13:08:17 2019 |
OttoRun | RUN | General | Cryo-3He | Run81 |
*************************
Run : 81
*************************
Start at 2019-06-29 13:08:04.442155
target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted
AGATA position +51-25 mm
TRIGGER :
- CMG
- M2
- VAMOS%1000
- CATS%1000
all X on MUVI disabled from trigger
all threshold of MUVI1 at 1 MeV to avoid as much as possible noise
LN filling problem solved !!!
pressure of the target slowly increasing about 820 mbar
We are filling the target to reach 1000 mbar
test cats1
Stop at 2019-06-29 19:22:32.016536
Stop comment: |
Fixed | e768s |
Thu Jul 18 10:44:52 2019 |
OttoRun | RUN | General | N/A | Run81 |
*************************
Run : 81
*************************
Start at 2019-07-18 04:21:23.813605
Time calibrator starting from MG3
strip by strip
Range 640 ns
Period 160 ns |
Fixed | e744s |
Thu Apr 11 17:43:44 2019 |
OttoRun | RUN | General | Empty | Run80 |
*************************
Run : 80
*************************
Start at 2019-04-11 17:42:44.178301
Run Conditions:
- Operator: Adrien
- Trigger: VAMOS %1
- Target/Source: hole
- Beam: 14O
- Intensity: ~ 5e+3
- Brho = 0.69186 Tm
- !CATS/VAMOS (CATS Efficiency)! : XXXX
Comment:
After removing division on Cats and Vamos.
Aligning beam, daq on for online monitoring
Stop at 2019-04-11 18:50:52.122268
Stop comment: Stop to remove target hole |
Fixed | e786s |
Sat Jun 29 12:07:56 2019 |
OttoRun | RUN | General | Cryo-3He | Run80 |
*************************
Run : 80
*************************
Start at 2019-06-29 12:07:36.615014
target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted
AGATA position +51-25 mm
TRIGGER :
- CMG
- M2
- VAMOS%1000
- CATS%1000
all X on MUVI disabled from trigger
all threshold of MUVI1 at 1 MeV to avoid as much as possible noise
LN filling problem solved !!!
pressure of the target slowly increasing about 820 mbar
We are filling the target to reach 1000 mbar
test cats1
Stop at 2019-06-29 13:03:26.602226
Stop comment: radioactive beam recovered |
Fixed | e768s |
Thu Jul 18 10:44:58 2019 |
OttoRun | RUN | General | N/A | Run80 |
*************************
Run : 80
*************************
Start at 2019-07-18 03:18:57.828894
Time calibrator starting from MG3
strip by strip
Range 640 ns
Period 160 ns |
Fixed | e768s |
Mon Jul 15 18:17:49 2019 |
Iulian, Serge | RUN | General | N/A | Run8 - TAC calibrations |
*************************
Run : 8
*************************
Start at 2019-07-15 18:17:21.942994
TAC Time calibrator
range 2.56 um
10 ns period |
Fixed | e744s |
Tue Apr 9 14:33:19 2019 |
OttoRun | RUN | General | N/A | Run8 |
*************************
Run : 8
*************************
Start at 2019-04-09 14:18:07.214936
Run Conditions:
- Operator: Adrien
- Trigger: MUST2
- Target/Source: none
- Beam: no
- Intensity:
Comment: Time Calibrator on MUGAST all detector using fsc
600ns on Y S1
Stop at 2019-04-09 14:33:20.112477
Stop comment: |
Fixed | TEST |
Fri Jun 21 17:14:55 2019 |
OttoRun | RUN | General | N/A | Run8 |
*************************
Run : 8
*************************
Start at 2019-06-21 17:14:12.777585
test
Stop at 2019-06-21 17:16:36.895632
Stop comment: |
Fixed | e744s |
Thu Apr 11 17:08:39 2019 |
OttoRun | RUN | General | Empty | Run79 |
*************************
Run : 79
*************************
Start at 2019-04-11 17:08:21.837538
Run Conditions:
- Operator: Adrien
- Trigger: VAMOS
- Target/Source: hole
- Beam: 14O
- Intensity: ~ 3e+3
- Brho = ?? Tm
- !CATS/VAMOS (CATS Efficiency)! : XXXX
Comment:
Aligning beam, daq on for online monitoring
Stop at 2019-04-11 17:31:48.540765
Stop comment: Entering to Remove CATS2 and remove VAMOS %1000 |
Fixed | e786s |
Sat Jun 29 09:32:52 2019 |
OttoRun | RUN | General | Cryo-3He | Run79 |
*************************
Run : 79
*************************
Start at 2019-06-29 09:31:41.957308
target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted
AGATA position +51-25 mm
TRIGGER :
- CMG
- M2
- VAMOS%1000
- CATS%1000
all X on MUVI disabled from trigger
all threshold of MUVI1 at 1 MeV to avoid as much as possible noise
LN filling problem solved !!!
pressure of the target slowly increasing about 800 mbar
We are filling the target to reach 1000 mbar
Stop at 2019-06-29 12:04:05.587989
Stop comment: |
Fixed | e768s |
Thu Jul 18 10:45:00 2019 |
OttoRun | RUN | General | N/A | Run79 |
*************************
Run : 79
*************************
Start at 2019-07-18 02:16:37.027826
Time calibrator starting from MG3
strip by strip
Range 640 ns
Period 160 ns
Stop at 2019-07-18 10:45:16.248349
Stop comment:
Stop at 2019-07-18 10:45:35.922721
Stop comment:
Stop at 2019-07-18 10:45:49.689858
Stop comment: |
Fixed | e744s |
Thu Apr 11 16:56:34 2019 |
OttoRun | RUN | General | Empty | Run78 |
*************************
Run : 78
*************************
Start at 2019-04-11 16:55:28.720017
Run Conditions:
- Operator: Adrien
- Trigger: Mugast
- Target/Source: hole
- Beam: 14O
- Intensity: ~ 3e+3
- Brho = ?? Tm
- !CATS/VAMOS (CATS Efficiency)! : XXXX
Comment:
Alignign beam, daq on for online monitoring
Stop at 2019-04-11 17:07:49.331973
Stop comment: changing trigger to vamos |
Fixed | e786s |
Sat Jun 29 06:18:06 2019 |
OttoRun | RUN | General | Cryo-3He | Run78 |
*************************
Run : 78
*************************
Start at 2019-06-29 06:17:46.891353
target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted
AGATA position +51-25 mm
TRIGGER :
- CMG
- M2
- VAMOS%1000
- CATS%1000
all X on MUVI disabled from trigger
all threshold of MUVI1 at 1 MeV to avoid as much as possible noise
beam back. Q pole issues still
Stop at 2019-06-29 08:57:15.767428
Stop comment:
HV of Drift Chamber in VAMOS reduced from 670V to 660V in order to avoid saturation, this will keep like that in the following. |
Fixed | e768s |
Thu Jul 18 10:44:53 2019 |
OttoRun | RUN | General | N/A | Run78 |
*************************
Run : 78
*************************
Start at 2019-07-18 01:14:14.943434
Time calibrator starting from MG3
strip by strip
Range 640 ns
Period 160 ns |
Fixed | e744s |
Thu Apr 11 08:55:23 2019 |
OttoRun | RUN | General | CH2 | Run77 |
*************************
Run : 77
*************************
Start at 2019-04-11 08:54:59.257928
Run Conditions:
- Operator: Nikola, Marion
- Trigger: Mugast & Vamos % 1000 & Cats2 % 40000
- Target/Source: CH2 105 um (Position 2)
- Beam: 14N
- Intensity: ~ 2.2e+5
- Brho = 0.437 Tm
- CATS 2 IN
Comment:
Stop at 2019-04-11 10:08:08.427024
Stop comment: Beam stopped for radiactive preparation |
Attachment 1: run77_tab0.png
|
|
Attachment 2: run77_tab2.png
|
|
Attachment 3: run77_tab3.png
|
|
Attachment 4: run77_tab4.png
|
|
Attachment 5: run77_tab5.png
|
|
Attachment 6: run77_tab6.png
|
|
Attachment 7: run77_time_calibrator.png
|
|
Attachment 8: run77_tab8.png
|
|
Attachment 9: run77_tab9.png
|
|
Attachment 10: run77_tab10.png
|
|
Attachment 11: run77_tab11.png
|
|
Attachment 12: run77_tab1.png
|
|
Attachment 13: Run77_1.png
|
|
Attachment 14: Run77_2.png
|
|
Attachment 15: Run77_3.png
|
|
Attachment 16: Run77_4.png
|
|
Attachment 17: Run77_5.png
|
|
Attachment 18: Run77_6.png
|
|
Attachment 19: Run77_7.png
|
|
Attachment 20: Run77_8.png
|
|
Attachment 21: Run77_9.png
|
|
Attachment 22: Run77_10.png
|
|
Attachment 23: Run77_11.png
|
|
Fixed | e786s |
Fri Jun 28 18:34:35 2019 |
OttoRun | RUN | General | Cryo-3He | Run77 |
========================================
*************************
Run : 77
*************************
Start at 2019-06-28 18:34:22.873862
target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted
AGATA position +51-25 mm
TRIGGER :
- CMG
- M2
- VAMOS%1000
- CATS%1000
all X on MUVI disabled from trigger
all threshold of MUVI1 at 1 MeV to avoid as much as possible noise
beam back. Q pole issues still
Stop at 2019-06-29 06:14:31.307098
Stop comment: |
Fixed | e786s |
Sat Jun 29 01:53:24 2019 |
OttoRun | RUN | General | Cryo-3He | Run77 |
OttoRun wrote: | *************************
Run : 77
*************************
Start at 2019-06-28 18:34:22.873862
target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted
AGATA position +51-25 mm
TRIGGER :
- CMG
- M2
- VAMOS%1000
- CATS%1000
all X on MUVI disabled from trigger
all threshold of MUVI1 at 1 MeV to avoid as much as possible noise
beam back. Q pole issues still |
Beam is back, again, at 1.53, weird behavior of CATS |
Attachment 1: 77CATS.png
|
|
Fixed | e786s |
Sat Jun 29 05:41:12 2019 |
OttoRun | RUN | General | Cryo-3He | Run77 |
OttoRun wrote: |
OttoRun wrote: | *************************
Run : 77
*************************
Start at 2019-06-28 18:34:22.873862
target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted
AGATA position +51-25 mm
TRIGGER :
- CMG
- M2
- VAMOS%1000
- CATS%1000
all X on MUVI disabled from trigger
all threshold of MUVI1 at 1 MeV to avoid as much as possible noise
beam back. Q pole issues still |
Beam is back, again, at 1.53, weird behavior of CATS |
|
Attachment 1: 78MM.png
|
|
Attachment 2: 80mmetof.png
|
|
Attachment 3: 79mg.png
|
|
Attachment 4: 81mmede.png
|
|
Attachment 5: 82mgetof.png
|
|
Attachment 6: tac.png
|
|
Attachment 7: corr.png
|
|
Attachment 8: cats.png
|
|