  MUGAST, Page 33 of 132  Not logged in MUGAST
Status ExpNbrdown Date Author Type Category Target-Source Subject
Fixede768s Sat Jul 20 21:10:57 2019 Faïrouz & SylvainRUNGeneralLiFRun 118 stopped because VXI Vamos turned offline because of overheat
Run 118 stopped because VXI Vamos turned offline of overheat
Fixede768s Sat Jul 20 23:18:02 2019 OttoRunRUNGeneralLiFRun119
*************************
Run : 119
*************************

Start at 2019-07-20 23:17:24.464827

Start after SECOND overtemperature of VXI VAMOS.

Stop at 2019-07-21 20:12:21.834156
Stop comment: stopped because intensity dropped to 6e6 pps (profiler) PCP working on optimization
Fixede768s Sat Jul 20 23:32:44 2019 Sylvain FairouzOFFLINEGeneral Trying to understand AGATA g-ray yields

Marlene, Nicolas, Christian, Fairouz wrote:
goal: we want to check whether the yields of the 1297.7 keV (from 19Ne 1536, 3/2+, l=1 state) and 1340 keV (from 19Ne 1615, 3/2-, l=2 state) g-ray lines is compatible with what we expect from x-sec calculations

conditions: we consider runs 112, 115 and 116

observations:
N(1297) = 17 g-rays
N(1340) = 23 g-rays

remarks:
Given x-sec from mirror nucleus (Francois 15N(7Limt) paper) we expect N(1340) >~ N(1297)

ingredients:
1. beam normalization
We use the TAC_FPMW_HF as a proxy. During the same time we integrate the prompt peak and compare the integrated OR VAMOS scaler, we find Scaler(OR VAMOS)/N(TAC) = 193.
We use the comparison of the OR VAMOS (pps) and the beam profiler (pps), we find profiler/OR VAMOS = 1e4
We obtain (2.17e5 + 6.44e4 + 3.52e5)*193*1e4 = 1.2 e12 15O incident particles (at profiler position!)

2. target thickness -> known to be 1.25 mg/cm2 of LiF (RBS measurement)
N(Li) = 1.25e-3 / (19+7) * 6.02e23 at/cm2

3.efficiencies
AGATA @ 1.3 MeV -> 7.5 %
VAMOS intrinsic -> 67 % (Eff_Check.C macro applied to run 116)
MUGAST -> 50 % (lab, geometry), needed because MUGAST is the main trigger

4. 19Ne charge state
we assume 40 % for 9+

5. g-ray branching ratio -> taken from NNDC

6. cross-section
we calculate differential cross section for the 1536-keV 19Ne state (FRESCO, FR-DWBA)
-> at 0 deg, ds/dW =
-> integrated x-sec in VAMOS acceptance (4 deg (lab) -> 20 deg (cm)) = 445 micro barn (C2Sa = 1)
-> taking into account know spectroscopic factor form analog state -> 89 micro barn

Conclusion:
for the 1536-keV 19Ne state we obtain 30 g-rays, instead of 17.

Discussion:
This factor of two could be well within the mirror symmetry assumption. However it seems it is also the case for the 1615-keV 19Ne state, so would not it be a systematic error with a factor of two missing somewhere (beam transmission between the profiler and the target? deficiencies?)?


We have investigated quickly the 15N(7Li,t)19F reaction based on run 0072. (gamma spectrum corresponding to all run data and with doppler correction with beta 0.08 can be found in /data/mugastX/mugast/acquisition/root/Run72_19Fe_8p.root)

To simplify the analysis the TS were restricted from 46000s to 52000s (/data/mugastX/mugast/acquisition/root/Run72_19Fe_8p_TSGate.root). The beam intentisty was assumed constant and equal to 1.e8 pps (for a total of 6.e11 beam like particle)

The 3/2+ and 3/2- states in 19F produce respectively 1356keV and 1348keV gamma. Counts were estimated roughly with double gaussian fit to be equal to
I(1348) = 21.9 counts
I(1256) = 16.4 counts

Assuming all others parameters are comparable with the previous analysis the counting obtained seems consistent with the measured Salpha=0,2 in previous experiment, Everything seems ok based on this very rough first analysis
Fixede768s Sat Jul 20 23:43:02 2019 mugastRUNGeneralLiFRun 119 (Problem with VXI fixed for the moment)
Run 119.
Or-Vamos= 1430 ---> I=1.43x10^7 pps
Attachment 1: scaler_run119_1.png
scaler_run119_1.png
Fixede768s Sat Jul 20 23:57:34 2019 DiegoOFFLINEGeneral VAMOS acceptance, comparison between 9+ and 10+ states
Figures above, simulation of VAMOS acceptance for 10+ and 9+ 19Ne charge states. Ratio between them quite big (0.58)
Figures below, comparison of the states with offline data. Ratio much smaller.
It seems like acceptance doesn't cut the 10+ charge state too much, but 10+ is less produced than 9+.
Attachment 1: VAMOS_acceptance.png
VAMOS_acceptance.png
Fixede768s Sun Jul 21 01:57:59 2019 Jenn & FranckONLINEGeneralLiFScreenshots run 119
Attachment 1: Diamond_r119_1.png
Diamond_r119_1.png
Attachment 2: Scalers_r119_1.png
Scalers_r119_1.png
Attachment 3: DCX_1d_r119_1.png
DCX_1d_r119_1.png
Attachment 4: IC_IC_r119_1.png
IC_IC_r119_1.png
Attachment 5: FP_r119_1.png
FP_r119_1.png
Attachment 6: Tmg_vamos_r119_1.png
Tmg_vamos_r119_1.png
Attachment 7: IC_r119_1.png
IC_r119_1.png
Attachment 8: MW_r119_1.png
MW_r119_1.png
Attachment 9: Ewire_r119_1.png
Ewire_r119_1.png
Attachment 10: Twire_r119_1.png
Twire_r119_1.png
Attachment 11: DCX_r119_1.png
DCX_r119_1.png
Attachment 12: mg_eraw_r119_1.png
mg_eraw_r119_1.png
Attachment 13: mg_timecal_r119_1.png
mg_timecal_r119_1.png
Attachment 14: ADC_r119_1.png
ADC_r119_1.png
Attachment 15: corr_r119_1.png
corr_r119_1.png
Attachment 16: TACs_r119_1.png
TACs_r119_1.png
Attachment 17: mg_etof_r119_1.png
mg_etof_r119_1.png
Attachment 18: mm_ede_r119_1.png
mm_ede_r119_1.png
Attachment 19: mm_etof_r119_1.png
mm_etof_r119_1.png
Attachment 20: mg_impact_r119_1.png
mg_impact_r119_1.png
Attachment 21: mm_impact_r119_1.png
mm_impact_r119_1.png
Fixede768s Sun Jul 21 04:48:52 2019 Jenn & FranckBEAMGeneralN/ABeam intensity check
OR Vamos: 1300 /s
Vamos gas profiler: 1.7x10^7 pps
Fixede768s Sun Jul 21 09:10:50 2019 Didier & NicolasBEAMGeneralN/ABeam intensity check

Jenn & Franck wrote:
OR Vamos: 1300 /s
Vamos gas profiler: 1.7x10^7 pps


OR VAMOS: ~ 1280 Hz
VAMOS gas profiler: 1.7e7 pps
Attachment 1: IMG_20190721_090616_beamcheck.jpg
IMG_20190721_090616_beamcheck.jpg
Fixede768s Sun Jul 21 10:58:39 2019 ChristianAGATAGeneralLiF13c resync
13c resync at 10.57
lost 5 min agata data
Fixede768s Sun Jul 21 11:49:59 2019 Christiqn qnd DidierRUNGeneralLiFRun 119 screenshots
Attachment 1: run_119_ICIC_MMETOF.png
run_119_ICIC_MMETOF.png
Attachment 2: run_119_scaler_grqph.png
run_119_scaler_grqph.png
Attachment 3: run_119_scalers.png
run_119_scalers.png
Fixede768s Sun Jul 21 14:54:28 2019 MarlèneRUNGeneral run 119 scalers
Attachment 1: run_119_scaler_beam2.png
run_119_scaler_beam2.png
Attachment 2: run_119_scaler_beam.png
run_119_scaler_beam.png
Fixede768s Sun Jul 21 16:39:38 2019 Marlène, AntoineBEAMGeneral IMPORTANT :How to send scalers to PCP
Go to PCP

Open a terminal
> ssh -X mugast@ganp154
passwd : Spiral1+

> ./LaunchScalers.sh
Fixede768s Sun Jul 21 16:42:20 2019 MarlèneOFFLINEGeneral Conversion of data from AGATA :update
When VAMOS VXI crashed (run 118) the Action file was recompiled.
As AGATA conversion is not copying the ACTION file for each run, it used the old version by default (from 16th of july).

We have modified the do_global_replay.sh so that it takes the ACTION file from the Replay directory where we have copied the ACTION file for run_0119.

So for conversion of file before run 119, the ACTION file in the zworkingDir of AGATA should be used.
For all conversion of runs after run 119, the ACTION file in the Replay directory should be used.
Fixede768s Sun Jul 21 17:02:25 2019 MarlèneBEAMGeneral beam intensity in the afternoon
Attachment 1: run_119_scaler3.png
run_119_scaler3.png
Fixede768s Sun Jul 21 17:45:54 2019 kseniiaAGATAGeneral Current total stats on 19Ne
Stats from runs 112, 115, 116, 117, 118, 119(~15h)
beta = 8%
Q = 9 & 10
Attachment 1: run112-119part-c.png
run112-119part-c.png
Attachment 2: run112-119part_zoom.png
run112-119part_zoom.png
Fixede768s Sun Jul 21 17:57:08 2019 ChristianOTHERGeneral Previous Papers - gamma decays from states in the 4-MeV region

In comment to previous entry (see thread), on the 2019 Hall paper on 19Ne transitions:

Note that Fig 4.c is described as: Gated on the 238-keV 5/2+ to ground-state transition. The 3897.5-keV transition from the 4141.8-keV state is observed for the first time. This means the spectrum is designed such that it wold not show the 4033 transition. Therefore, whether they are sensitive to this is an open question. See also PhD thesis of Hall for further details?
Fixede768s Sun Jul 21 17:59:04 2019 ChristianVAMOSHardwareLiFDiamond detector trigger level and rate dependency
The diamond detector has tended to increase in rate when it is hot (peaking late afternoon) and decreasing when it is cold (with a minimum just before sun-rise). Some of the rate could therefore be temperature dependent noise levels rising above the trigger threshold.

We are currently triggering around 70mV, far below the peak signal value of 1.4V (by about a factor of 20). While we cannot expect to see most of them at the peak value because of poor resolution and potentially different particle energies, we should expect the vast majority of real heavy-ion signals to fall above 1/4 of the peak height. When we go into the vault on Tuesday, we should therefore raise the thresholds by about a factor of 5, to see whether this stabilises the rate.

Irrespective of this, with the beam deposited so far from the detector as is the case, we should only see halo and/or reaction products, and we would not be able to use this as a beam intensity monitor, irrespective of performance. At best as a beam-quality monitor.
Attachment 1: IMG_1065.JPG
IMG_1065.JPG
Attachment 2: IMG_1064.JPG
IMG_1064.JPG
Fixede768s Sun Jul 21 20:13:57 2019 Kseniia, Franco & NicolasBEAMGeneralN/ABeam intensity check

Didier & Nicolas wrote:

Jenn & Franck wrote:
OR Vamos: 1300 /s
Vamos gas profiler: 1.7x10^7 pps


OR VAMOS: ~ 1280 Hz
VAMOS gas profiler: 1.7e7 pps


OR VAMOS: ~ 1700 Hz
VAMOS gas profiler: 6e6 pps

beam moved in the horizontal plane (compare last 2 pictures of beam profiler)
Attachment 1: IMG_20190721_194258_beamcheck.jpg
IMG_20190721_194258_beamcheck.jpg
Fixede768s Sun Jul 21 21:22:58 2019 OttoRunRUNGeneralLiFRun120
*************************
Run : 120
*************************

Start at 2019-07-21 21:19:46.530036

beam intensity after optimization 1.6e7 pps
OR VAMOS ~ 1500 Hz

Target: 1.25 mg/cm2 (nb 1) 7LiF

Triggers
MUGAST (backward, MM, 90 deg)
VAMOS %100
BaF2 %100

Stop at 2019-07-21 21:52:56.667021
Stop comment:
Fixede768s Sun Jul 21 22:17:18 2019 OttoRunRUNGeneralLiFRun121
*************************
Run : 121
*************************

Start at 2019-07-21 21:54:19.075515

beam intensity after optimization 1.6e7 pps
OR VAMOS ~ 1500 Hz

Target: 1.25 mg/cm2 (nb 1) 7LiF

Triggers
MUGAST (backward, MM, 90 deg)
VAMOS %100
BaF2 %100

Stop at 2019-07-22 05:09:23.310656
Stop comment: Problems with AGATA. Need to stop acquisition to recover AGATA validation