Status |
ExpNbr |
Date |
Author |
Type |
Category |
Target-Source |
Subject |
Fixed | e744s |
Thu Apr 11 00:06:50 2019 |
A. Lemasson | RUN | General | N/A | Run58 |
Without Finger 21 kHz -> Reduced to 800 Hz with finger |
Attachment 1: XfTFFinger.png
|
|
Attachment 2: run58_Brho_Cats2_14N.png
|
|
Fixed | e744s |
Thu Apr 11 00:23:21 2019 |
Iulian François Marlène | RUN | Hardware | CH2 | TAC changes & target |
We have inserted the 2nd target CH2 104 um.
We have changed the range of the TAC MUGAST-HF from 1us to 200ns.
TAC VAMOS-CATS1 (in fact CATS2) was not working in the previous run because there was no delay on CATS. We send now CATS DELAY x2, meaning that the delay is 300ns. |
Fixed | e744s |
Thu Apr 11 00:28:44 2019 |
OttoRun | RUN | General | CH2 | Run59 |
*************************
Run : 59
*************************
Start at 2019-04-11 00:27:28.387554
Run Conditions:
- Operator: A. Lemasson
- Trigger: VAMOS
- Target/Source: CH2 100 um (Position 2)
- Beam: 14N
- Intensity: ~ 2e+4
- Brho = 0.79067 Tm
- CATS 2 IN
Comment:
Target inserted (100 um ) and Cats2 in
Placing Finger on The beam
Stop at 2019-04-11 00:49:00.966603
Stop comment: |
Attachment 1: run59_Brho_Cats2_15N6_targetch2.png
|
|
Fixed | TEST |
Thu Apr 11 00:38:13 2019 |
OttoH.V. | OTHER | Hardware | N/A | HVM_ERROR |
*** hvm watcher error ***
FP_DC2_w2 Slot: 2 (A1821HMP) Channel:3 Vmon: 674.75 V Imon: 0.002 uA Status: OVER CURRENT
|
Fixed | TEST |
Thu Apr 11 00:40:33 2019 |
OttoH.V. | OTHER | Hardware | N/A | HVM_ERROR |
*** hvm watcher error ***
FP_DC2_w2 Slot: 2 (A1821HMP) Channel:3 Vmon: 8 V Imon: 0 uA Status: INTERNAL TRIP
|
Fixed | TEST |
Thu Apr 11 00:57:07 2019 |
OttoH.V. | OTHER | Hardware | N/A | HVM_ERROR |
*** hvm watcher error ***
FP_CATS3 Slot: 2 (A1821HMP) Channel:9 Vmon: 559.5 V Imon: 0.035 uA Status: OVER CURRENT
|
Fixed | e744s |
Thu Apr 11 00:57:34 2019 |
iulian | OTHER | Hardware | CH2 | CH2 target thickness |
target CH2 104 um (2nd position)
Brho 14N6+ CATS2 |
Fixed | TEST |
Thu Apr 11 00:57:58 2019 |
OttoH.V. | OTHER | Hardware | N/A | HVM_ERROR |
*** hvm watcher error ***
FP_CATS4 Slot: 2 (A1821HMP) Channel:10 Vmon: 524.5 V Imon: 0.94 uA Status: OVER CURRENT
|
Fixed | TEST |
Thu Apr 11 00:58:05 2019 |
OttoH.V. | OTHER | Hardware | N/A | HVM_ERROR |
*** hvm watcher error ***
FP_CATS3 Slot: 2 (A1821HMP) Channel:9 Vmon: 53 V Imon: 0 uA Status: INTERNAL TRIP
|
Fixed | e744s |
Thu Apr 11 01:00:46 2019 |
OttoRun | RUN | General | CH2 | Run60 |
*************************
Run : 60
*************************
Start at 2019-04-11 00:59:35.983897
Run Conditions:
- Operator: A. Lemasson
- Trigger: VAMOS
- Target/Source: CH2 100 um (Position 3)
- Beam: 14N
- Intensity: ~ 2e+4
- Brho = 0.79067 Tm
- CATS 2 IN
Comment:
Target inserted (100 um ) and Cats2 in
Placing Finger on The beam
Vamos IC10 placed in IC12 electronics
|
Fixed | e744s |
Thu Apr 11 01:05:35 2019 |
OttoRun | RUN | General | CH2 | Run61 |
*************************
Run : 61
*************************
Start at 2019-04-11 01:04:59.561319
Run Conditions:
- Operator: A. Lemasson
- Trigger: VAMOS
- Target/Source: CH2 100 um (Position 3)
- Beam: 14N
- Intensity: ~ 2e+4
- Brho = 0.79067 Tm
- CATS 2 IN
Comment:
Same as run 60 after DAQ NARVAL crash. (broken link 102 ???)
Target inserted (100 um ) and Cats2 in
Placing Finger on The beam
Vamos IC10 placed in IC12 electronics
|
Fixed | e744s |
Thu Apr 11 01:09:05 2019 |
OttoRun | RUN | General | CH2 | Run62 |
*************************
Run : 62
*************************
Start at 2019-04-11 01:08:38.422386
Run Conditions:
- Operator: A. Lemasson
- Trigger: VAMOS
- Target/Source: CH2 100 um (Position 3)
- Beam: 14N
- Intensity: ~ 2e+4
- Brho = 0.79067 Tm
- CATS 2 IN
Comment:
Same as run 60 after DAQ NARVAL crash. (DC not sending Data, Kill Narval on ganp507 and vamosng)
Target inserted (100 um ) and Cats2 in
Placing Finger on The beam
Vamos IC10 placed in IC12 electronics
|
Fixed | e744s |
Thu Apr 11 01:16:30 2019 |
OttoRun | RUN | General | CH2 | Run63 |
*************************
Run : 63
*************************
Start at 2019-04-11 01:16:17.572487
Run Conditions:
- Operator: A. Lemasson
- Trigger: VAMOS
- Target/Source: CH2 100 um (Position 3)
- Beam: 14N
- Intensity: ~ 2e+4
- Brho = 0.79067 Tm
- CATS 2 IN
Comment:
Same as run 60 after DAQ NARVAL crash. (DC not sending Data, Kill Narval on ganp507 and vamosng)
AGAIN After Crash
|
Fixed | TEST |
Thu Apr 11 01:16:38 2019 |
OttoH.V. | OTHER | Hardware | N/A | HVM_ERROR |
*** hvm watcher error ***
FP_CATS3 Slot: 2 (A1821HMP) Channel:9 Vmon: 0 V Imon: 0 uA Status: INTERNAL TRIP
FP_CATS4 Slot: 2 (A1821HMP) Channel:10 Vmon: 0 V Imon: 0.06 uA Status: INTERNAL TRIP
|
Fixed | e744s |
Thu Apr 11 01:21:39 2019 |
OttoRun | RUN | General | CH2 | Run64 |
*************************
Run : 64
*************************
Start at 2019-04-11 01:21:02.741720
Run Conditions:
- Operator: A. Lemasson
- Trigger: VAMOS
- Target/Source: CH2 100 um (Position 3)
- Beam: 14N
- Intensity: ~ 2e+4
- Brho = 0.79067 Tm
- CATS 2 IN
Comment:
Data OK for Reference Target Thickness measurement
Same as run 60 after DAQ NARVAL crash. (DC not sending Data, Kill Narval on ganp507 and vamosng)
AGAIN After Crash CRASH in NUMEXO 63 - BI STORAGE error
(01:22:48 - 11/04/2019) FATAL - log from NARVAL.NUMEXO2_61 - Bloc_Catcher.T_Communication : requested size too high available in actor 1024 requested in header 5376 do not use this actor
(01:22:48 - 11/04/2019) FATAL - log from NARVAL.NUMEXO2_61 - NUMEXO2_61 put sub system NARVAL in error with message :Bloc_Catcher.T_Communication : actor buffer size too small
(01:22:48 - 11/04/2019) ERROR - log from NARVAL.NUMEXO2_61 - Bloc_Catcher.T_Communication : BlC : exception,STORAGE_ERROR,s-intman.adb:136 explicit raise
(01:22:48 - 11/04/2019) ERROR - log from NARVAL.NUMEXO2_61 - Bloc_Catcher.T_Communication : BPR2 : exception,STORAGE_ERROR,s-intman.adb:136 explicit raise
(01:22:48 - 11/04/2019) ERROR - log from NARVAL.NUMEXO2_61 - Bloc_Catcher.T_Communication : BPR ,STORAGE_ERROR,s-intman.adb:136 explicit raise
(01:22:48 - 11/04/2019) ERROR - log from NARVAL.NUMEXO2_61 - Bloc_Catcher.T_Communication : exception ,STORAGE_ERROR,s-intman.adb:136 explicit raise
|
Fixed | e744s |
Thu Apr 11 01:37:57 2019 |
OttoRun | RUN | General | CH2 | Run65 |
*************************
Run : 65
*************************
Start at 2019-04-11 01:37:16.190996
Run Conditions:
- Operator: A. Lemasson
- Trigger: VAMOS
- Target/Source: C 73 um (Position 4)
- Beam: 14N
- Intensity: ~ 2e+4
- Brho = 0.79067 Tm
- CATS 2 IN
Comment:
Changed target to C 73 um |
Fixed | e744s |
Thu Apr 11 01:48:40 2019 |
OttoRun | RUN | General | Carbon | Run66 |
*************************
Run : 66
*************************
Start at 2019-04-11 01:48:23.997955
Run Conditions:
- Operator: A. Lemasson
- Trigger: VAMOS
- Target/Source: C 73 um (Position 4)
- Beam: 14N
- Intensity: ~ 2e+4
- Brho = 0.79067 Tm
- CATS 2 IN
Comment:
Changed target to C 73 um
Stop at 2019-04-11 01:58:13.528779
Stop comment: |
Fixed | TEST |
Thu Apr 11 01:56:59 2019 |
OttoH.V. | OTHER | Hardware | N/A | HVM_ERROR |
*** hvm watcher error ***
FP_DC2_w2 Slot: 2 (A1821HMP) Channel:3 Vmon: 672.5 V Imon: 0.065 uA Status: OVER CURRENT
|
Fixed | TEST |
Thu Apr 11 01:59:19 2019 |
OttoH.V. | OTHER | Hardware | N/A | HVM_ERROR |
*** hvm watcher error ***
FP_DC2_w2 Slot: 2 (A1821HMP) Channel:3 Vmon: 9.25 V Imon: 0 uA Status: INTERNAL TRIP
|
Fixed | e744s |
Thu Apr 11 02:09:12 2019 |
OttoRun | RUN | General | CH2 | Run67 |
*************************
Run : 67
*************************
Start at 2019-04-11 02:08:39.018915
Run Conditions:
- Operator: A. Lemasson
- Trigger: VAMOS
- Target/Source: CH2 73 um (Position 5)
- Beam: 14N
- Intensity: ~ 2e+4
- Brho = 0.79067 Tm
- CATS 2 IN
Comment:
Changed target to CH2 73 um
Stop at 2019-04-11 02:13:21.475620
Stop comment: |