MUST2
SAMURAI_2018
S3-LEB-LPC
SUPERNEMO
MUGAST
EXPAND
SCALP
GALATRON
HiCARI
VELO
MUGAST, Page 1 of 132
Not logged in
Find
|
Login
Full
| Summary |
Threaded
2636 Entries
Goto page 1,
2
,
3
...
130
,
131
,
132
Next
Status
ExpNbr
Date
Author
Type
Category
Target-Source
Subject
Text
Fixed
e793s
Tue Mar 10 18:46:13 2020
Franco & Marlène
GRIT
Hardware
Changes in MUGAST
- MG6 (FZ) is disconnected.
- MG7 (500um) is dismounted and MG8 (375um)
is replacing it.
Fixed
e793s
Wed Mar 11 19:02:54 2020
mugast
GRIT
General
3-alpha
Alpha calibration for e793s - run 600
========================================
========================================
========================================
Fixed
e793s
Wed Mar 11 22:01:18 2020
mugast
GRIT
General
3-alpha
Alpha calibration - run 600
After 2h30' of run we checked the temperatures
of the boards and we noticed that MG7 had
63 degrees on both X and Y, which is very
Fixed
e793s
Thu Mar 12 14:54:13 2020
Marlène & Franco
GRIT
General
3-alpha
Run 0601 - alpha source calibration
We start the alpha source calibration run
with pedestal suppression (except for MG7
and MG9) and cooling at 0 deg.
Fixed
e793s
Thu Mar 12 15:48:58 2020
Franco&Marlène
GRIT
General
Summary and TO DO list 12/03/2020
What remains to be done
- Do the time calibration of all boards (time
calibrator has to be connected)
Fixed
e793s
Thu Mar 12 17:25:19 2020
mugast
GRIT
General
3-alpha
run 0601 - alpha source with pedestals
Alpha source calibration run without pedestal
suppression and cooling at 0 deg.
Fixed
e793s
Fri Jan 29 15:11:54 2021
Adrien
VACUUM
General
N/A
MUGAST undervacuum and powered up
MUGAST is undervacuum thanks to Arnaud and
Laurent.
Fixed
e793s
Mon Feb 8 10:27:23 2021
OttoTime
RUN
General
N/A
TimeCalibratorRun
*** Run Time Calibrator FSC ***
- Run: 605
- MUVI: MUVI1
Fixed
e793s
Mon Feb 8 10:39:35 2021
OttoTime
RUN
General
N/A
TimeCalibratorRun
*** Run Time Calibrator FSC ***
- Run: 606
- MUVI: MUVI1
Fixed
e793s
Mon Feb 8 10:51:48 2021
OttoTime
RUN
General
N/A
TimeCalibratorRun
*** Run Time Calibrator FSC ***
- Run: 607
- MUVI: MUVI2
Fixed
e793s
Mon Feb 8 11:04:04 2021
OttoTime
RUN
General
N/A
TimeCalibratorRun
*** Run Time Calibrator FSC ***
- Run: 608
- MUVI: MUVI2
Fixed
e793s
Mon Feb 8 11:16:18 2021
OttoTime
RUN
General
N/A
TimeCalibratorRun
*** Run Time Calibrator FSC ***
- Run: 609
- MUVI: MUVI3
Fixed
e793s
Mon Feb 8 11:25:29 2021
Adrien and Freddy
S1
General
N/A
FSC run
we ran FSC on all card. S1 is dismounted.
Pulser goes on all channel and works well
so electronic is fine.
Fixed
e793s
Mon Feb 8 11:28:38 2021
OttoTime
RUN
General
N/A
TimeCalibratorRun
*** Run Time Calibrator FSC ***
- Run: 610
- MUVI: MUVI3
Fixed
e793s
Mon Feb 8 16:20:42 2021
OttoTime
RUN
General
N/A
TimeCalibratorRun
*** Run Time Calibrator FSC ***
- Run: 611
- MUVI: MUVI2
Fixed
e793s
Mon Feb 8 16:33:22 2021
OttoTime
RUN
General
N/A
TimeCalibratorRun
*** Run Time Calibrator FSC ***
- Run: 612
- MUVI: MUVI2
Fixed
e793s
Fri Feb 12 08:52:30 2021
mugast
RUN
General
3-alpha
run 613: long alpha run
all Muvi in trigger, mugast stand alone. all
detector Bias on.
Fixed
e793s
Fri Feb 12 18:07:24 2021
OttoTime
RUN
General
N/A
TimeCalibratorRun
*** Run Time Calibrator FSC ***
- Run: 614
- MUVI: MUVI2
Fixed
e793s
Fri Feb 12 18:12:00 2021
Adrien
S1
Hardware
N/A
Kapton change
Arnaud open today and change the Y Kapton
(YA5) of the S1. Everything under vacuum/cooled
again so I ran FSC, hopefully this will fix
Fixed
e793s
Fri Feb 12 18:19:52 2021
OttoTime
RUN
General
N/A
TimeCalibratorRun
*** Run Time Calibrator FSC ***
- Run: 615
- MUVI: MUVI2
Goto page 1,
2
,
3
...
130
,
131
,
132
Next
ELOG Home Page