MUST2 SAMURAI_2018 S3-LEB-LPC SUPERNEMO MUGAST EXPAND SCALP GALATRON HiCARI VELO
  MUGAST, Page 1 of 132  Not logged in MUGAST
Status ExpNbrdown Date Author Type Category Target-Source Subject Text Attachments
Fixede793s Tue Mar 10 18:46:13 2020 Franco & MarlèneGRITHardware Changes in MUGAST- MG6 (FZ) is disconnected.
- MG7 (500um) is dismounted and MG8 (375um)
is replacing it.
 MG_X_raw_11032020.pngMG_Y_raw_11032020.png 
Fixede793s Wed Mar 11 19:02:54 2020 mugastGRITGeneral3-alphaAlpha calibration for e793s - run 600========================================
========================================
========================================
  
Fixede793s Wed Mar 11 22:01:18 2020 mugastGRITGeneral3-alphaAlpha calibration - run 600After 2h30' of run we checked the temperatures
of the boards and we noticed that MG7 had
63 degrees on both X and Y, which is very
 MG7X_run600_Eevolution.png 
Fixede793s Thu Mar 12 14:54:13 2020 Marlène & FrancoGRITGeneral3-alphaRun 0601 - alpha source calibrationWe start the alpha source calibration run
with pedestal suppression (except for MG7
and MG9) and cooling at 0 deg.
 HV_run601.png 
Fixede793s Thu Mar 12 15:48:58 2020 Franco&MarlèneGRITGeneral Summary and TO DO list 12/03/2020What remains to be done
- Do the time calibration of all boards (time
calibrator has to be connected)
  
Fixede793s Thu Mar 12 17:25:19 2020 mugastGRITGeneral3-alpharun 0601 - alpha source with pedestalsAlpha source calibration run without pedestal
suppression and cooling at 0 deg.
  
Fixede793s Fri Jan 29 15:11:54 2021 AdrienVACUUMGeneralN/AMUGAST undervacuum and powered upMUGAST is undervacuum thanks to Arnaud and
Laurent. 
  
Fixede793s Mon Feb 8 10:27:23 2021 OttoTimeRUNGeneralN/ATimeCalibratorRun*** Run Time Calibrator FSC ***
  - Run: 605
  - MUVI: MUVI1
  
Fixede793s Mon Feb 8 10:39:35 2021 OttoTimeRUNGeneralN/ATimeCalibratorRun*** Run Time Calibrator FSC ***
  - Run: 606
  - MUVI: MUVI1
  
Fixede793s Mon Feb 8 10:51:48 2021 OttoTimeRUNGeneralN/ATimeCalibratorRun*** Run Time Calibrator FSC ***
  - Run: 607
  - MUVI: MUVI2
  
Fixede793s Mon Feb 8 11:04:04 2021 OttoTimeRUNGeneralN/ATimeCalibratorRun*** Run Time Calibrator FSC ***
  - Run: 608
  - MUVI: MUVI2
  
Fixede793s Mon Feb 8 11:16:18 2021 OttoTimeRUNGeneralN/ATimeCalibratorRun*** Run Time Calibrator FSC ***
  - Run: 609
  - MUVI: MUVI3
  
Fixede793s Mon Feb 8 11:25:29 2021 Adrien and FreddyS1GeneralN/AFSC runwe ran  FSC on all card. S1 is dismounted.
Pulser goes on all channel and works well
so electronic is fine.
 fscE_08022023.pngfscT_08022022.png 
Fixede793s Mon Feb 8 11:28:38 2021 OttoTimeRUNGeneralN/ATimeCalibratorRun*** Run Time Calibrator FSC ***
  - Run: 610
  - MUVI: MUVI3
  
Fixede793s Mon Feb 8 16:20:42 2021 OttoTimeRUNGeneralN/ATimeCalibratorRun*** Run Time Calibrator FSC ***
  - Run: 611
  - MUVI: MUVI2
  
Fixede793s Mon Feb 8 16:33:22 2021 OttoTimeRUNGeneralN/ATimeCalibratorRun*** Run Time Calibrator FSC ***
  - Run: 612
  - MUVI: MUVI2
  
Fixede793s Fri Feb 12 08:52:30 2021 mugastRUNGeneral3-alpharun 613: long alpha run all Muvi in trigger, mugast stand alone. all
detector Bias on.
  
Fixede793s Fri Feb 12 18:07:24 2021 OttoTimeRUNGeneralN/ATimeCalibratorRun*** Run Time Calibrator FSC ***
  - Run: 614
  - MUVI: MUVI2
  
Fixede793s Fri Feb 12 18:12:00 2021 AdrienS1HardwareN/AKapton changeArnaud open today and change the Y Kapton
(YA5) of the S1. Everything under vacuum/cooled
again so I ran FSC, hopefully this will fix
  
Fixede793s Fri Feb 12 18:19:52 2021 OttoTimeRUNGeneralN/ATimeCalibratorRun*** Run Time Calibrator FSC ***
  - Run: 615
  - MUVI: MUVI2
  

ELOG Home Page