MUST2 SAMURAI_2018 S3-LEB-LPC SUPERNEMO MUGAST EXPAND SCALP GALATRON HiCARI VELO
  MUGAST, Page 18 of 132  Not logged in MUGAST
Status ExpNbrdown Date Author Type Category Target-Source Subject
Fixede786s Thu Jun 27 17:59:20 2019 OttoRunRUNGeneralCryo-3HeRun70
*************************
Run : 70
*************************

Start at 2019-06-27 17:58:29.855367

target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted

AGATA position +51-25 mm

TRIGGER :
- CMG
- M2
- VAMOS%1000
- CATS%1000

back to normal

Stop at 2019-06-27 18:37:51.816766
Stop comment: type your message here
Fixede786s Thu Jun 27 18:48:51 2019 OttoRunRUNGeneralCryo-3HeRun71
*************************
Run : 71
*************************

Start at 2019-06-27 18:47:56.171341

target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted

AGATA position +51-25 mm

TRIGGER :
- MG
- square (not biased)
- M2


all X on MUVI disabled from trigger
all threshold of MUVI1 at 1 MeV to avoid as much as possible noise

Stop at 2019-06-27 20:23:56.630240
Stop comment:
Fixede786s Thu Jun 27 20:26:27 2019 OttoRunRUNGeneralCryo-3HeRun73
*************************
Run : 73
*************************

Start at 2019-06-27 20:26:17.916314

target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted

AGATA position +51-25 mm

TRIGGER :
- CMG
- M2
- VAMOS%1000
- CATS%1000

all X on MUVI disabled from trigger
all threshold of MUVI1 at 1 MeV to avoid as much as possible noise

There is no run 72

Stop at 2019-06-27 23:04:45.884452
Stop comment: AGATA DAQ is rejectting all events
Fixede786s Thu Jun 27 23:28:30 2019 OttoRunRUNGeneralCryo-3HeRun74
*************************
Run : 74
*************************

Start at 2019-06-27 23:25:00.455575

target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted

AGATA position +51-25 mm

TRIGGER :
- CMG
- M2
- VAMOS%1000
- CATS%1000

all X on MUVI disabled from trigger
all threshold of MUVI1 at 1 MeV to avoid as much as possible noise

Stop at 2019-06-28 09:31:40.775271
Stop comment:
Fixede786s Fri Jun 28 04:21:48 2019 Daniele MengoniRUNGeneralCryo-3Herun 74
screenshots run 74 attached

frequent error message on the FP DC Wi1 --> defqlt. The elog said so, card tripped, but spectra are still present .. puzzling
Attachment 1: Q_vs_Xf_K1.png
Q_vs_Xf_K1.png
Attachment 2: Q_vs_Xf_K2.png
Q_vs_Xf_K2.png
Attachment 3: Q_vs_Xf_K3.png
Q_vs_Xf_K3.png
Attachment 4: Q_vs_Xf_K4.png
Q_vs_Xf_K4.png
Attachment 5: Q_vs_Xf_K5.png
Q_vs_Xf_K5.png
Attachment 6: Q_vs_Xf_K6.png
Q_vs_Xf_K6.png
Attachment 7: Q_vs_Xf_K7.png
Q_vs_Xf_K7.png
Attachment 8: Q_vs_Xf_K8.png
Q_vs_Xf_K8.png
Attachment 9: Q_vs_Xf_K9.png
Q_vs_Xf_K9.png
Attachment 10: Q_vs_Xf_K10.png
Q_vs_Xf_K10.png
Attachment 11: Q_vs_Xf_K11.png
Q_vs_Xf_K11.png
Attachment 12: Q_vs_Xf_K12.png
Q_vs_Xf_K12.png
Attachment 13: Q_vs_Xf_K13.png
Q_vs_Xf_K13.png
Attachment 14: Q_vs_Xf_K14.png
Q_vs_Xf_K14.png
Attachment 15: Q_vs_Xf_K15.png
Q_vs_Xf_K15.png
Attachment 16: Q_vs_Xf_K16.png
Q_vs_Xf_K16.png
Attachment 17: Q_vs_Xf_K17.png
Q_vs_Xf_K17.png
Attachment 18: Q_vs_Xf_K18.png
Q_vs_Xf_K18.png
Fixede786s Fri Jun 28 08:54:50 2019 FreddyRUNGeneralCryo-3Herun 74

Daniele Mengoni wrote:
screenshots run 74 attached

frequent error message on the FP DC Wi1 --> defqlt. The elog said so, card tripped, but spectra are still present .. puzzling




Put the voltage back at 8h00 when i arrived.
Fixede786s Fri Jun 28 08:57:21 2019 FreddyCATSGeneralCryo-3HeCATS scaler not compatible with beam rate
The CATS scaler si now around 100-120 kHz while the beam rate is still 40 kHz.

I checked the CATS signal and there is clearly noise who induce double logic signals. Not all of them are validated by FAG.

We should raise the threshold- call sebastien

9h32 : Stop beam to enter and raise the threshold.
Attachment 1: ScalerCATSproblem.png
ScalerCATSproblem.png
Fixede786s Fri Jun 28 09:56:59 2019 Bertrand, Valerian and AdrienBEAMGeneralN/ADipole stopped in the morning?
Dipole restart at 10. Shoudl be checked with time stamp.
Fixede786s Fri Jun 28 10:00:25 2019 OttoRunRUNGeneralCryo-3HeRun75
*************************
Run : 75
*************************

Start at 2019-06-28 09:59:59.505010

target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted

AGATA position +51-25 mm

TRIGGER :
- CMG
- M2
- VAMOS%1000
- CATS%1000

all X on MUVI disabled from trigger
all threshold of MUVI1 at 1 MeV to avoid as much as possible noise

CATS threshold raised

Stop at 2019-06-28 14:33:09.843668
Stop comment:
Fixede786s Fri Jun 28 15:25:12 2019 OttoRunRUNGeneralCryo-3HeRun76
*************************
Run : 76
*************************

Start at 2019-06-28 15:24:51.867946

target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted

AGATA position +51-25 mm

TRIGGER :
- CMG
- M2
- VAMOS%1000
- CATS%1000

all X on MUVI disabled from trigger
all threshold of MUVI1 at 1 MeV to avoid as much as possible noise

beam back. Q pole issues still

Stop at 2019-06-28 16:55:27.258794
Stop comment:
Fixede786s Fri Jun 28 18:34:35 2019 OttoRunRUNGeneralCryo-3HeRun77
========================================
*************************
Run : 77
*************************

Start at 2019-06-28 18:34:22.873862

target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted

AGATA position +51-25 mm

TRIGGER :
- CMG
- M2
- VAMOS%1000
- CATS%1000

all X on MUVI disabled from trigger
all threshold of MUVI1 at 1 MeV to avoid as much as possible noise

beam back. Q pole issues still

Stop at 2019-06-29 06:14:31.307098
Stop comment:
Fixede786s Sat Jun 29 01:53:24 2019 OttoRunRUNGeneralCryo-3HeRun77

OttoRun wrote:
*************************
Run : 77
*************************

Start at 2019-06-28 18:34:22.873862

target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted

AGATA position +51-25 mm

TRIGGER :
- CMG
- M2
- VAMOS%1000
- CATS%1000

all X on MUVI disabled from trigger
all threshold of MUVI1 at 1 MeV to avoid as much as possible noise

beam back. Q pole issues still

Beam is back, again, at 1.53, weird behavior of CATS
Attachment 1: 77CATS.png
77CATS.png
Fixede786s Sat Jun 29 05:41:12 2019 OttoRunRUNGeneralCryo-3HeRun77

OttoRun wrote:

OttoRun wrote:
*************************
Run : 77
*************************

Start at 2019-06-28 18:34:22.873862

target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted

AGATA position +51-25 mm

TRIGGER :
- CMG
- M2
- VAMOS%1000
- CATS%1000

all X on MUVI disabled from trigger
all threshold of MUVI1 at 1 MeV to avoid as much as possible noise

beam back. Q pole issues still

Beam is back, again, at 1.53, weird behavior of CATS
Attachment 1: 78MM.png
78MM.png
Attachment 2: 80mmetof.png
80mmetof.png
Attachment 3: 79mg.png
79mg.png
Attachment 4: 81mmede.png
81mmede.png
Attachment 5: 82mgetof.png
82mgetof.png
Attachment 6: tac.png
tac.png
Attachment 7: corr.png
corr.png
Attachment 8: cats.png
cats.png
Fixede786s Sat Jun 29 05:56:40 2019 OttoRunRUNGeneralCryo-3HeRun77

OttoRun wrote:

OttoRun wrote:

OttoRun wrote:
*************************
Run : 77
*************************

Start at 2019-06-28 18:34:22.873862

target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted

AGATA position +51-25 mm

TRIGGER :
- CMG
- M2
- VAMOS%1000
- CATS%1000

all X on MUVI disabled from trigger
all threshold of MUVI1 at 1 MeV to avoid as much as possible noise

beam back. Q pole issues still

Beam is back, again, at 1.53, weird behavior of CATS


After beam is back we found 350Hz on vamos focal plane, with 20kHz current. After Brho scanning we found a Brho distribution going from 1Tm to 0.4Tm. Observation of gamma rays indicates no accumulation of radioactivity in the chamber with respect to previous runs. We conclude that we may have ice layers up to 500um in the target but very in-homogeneous since we also observe 1Tm Brho. To verify maybe increase target pressure to modify the structure of the ice. CATS is tripping continuously. No VACUUM changes observed.
Fixede786s Sat Jun 29 06:18:06 2019 OttoRunRUNGeneralCryo-3HeRun78
*************************
Run : 78
*************************

Start at 2019-06-29 06:17:46.891353

target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted

AGATA position +51-25 mm

TRIGGER :
- CMG
- M2
- VAMOS%1000
- CATS%1000

all X on MUVI disabled from trigger
all threshold of MUVI1 at 1 MeV to avoid as much as possible noise

beam back. Q pole issues still

Stop at 2019-06-29 08:57:15.767428
Stop comment:

HV of Drift Chamber in VAMOS reduced from 670V to 660V in order to avoid saturation, this will keep like that in the following.
Fixede786s Sat Jun 29 09:32:52 2019 OttoRunRUNGeneralCryo-3HeRun79
*************************
Run : 79
*************************

Start at 2019-06-29 09:31:41.957308

target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted

AGATA position +51-25 mm

TRIGGER :
- CMG
- M2
- VAMOS%1000
- CATS%1000

all X on MUVI disabled from trigger
all threshold of MUVI1 at 1 MeV to avoid as much as possible noise

LN filling problem solved !!!
pressure of the target slowly increasing about 800 mbar
We are filling the target to reach 1000 mbar

Stop at 2019-06-29 12:04:05.587989
Stop comment:
Fixede786s Sat Jun 29 12:07:56 2019 OttoRunRUNGeneralCryo-3HeRun80
*************************
Run : 80
*************************

Start at 2019-06-29 12:07:36.615014

target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted

AGATA position +51-25 mm

TRIGGER :
- CMG
- M2
- VAMOS%1000
- CATS%1000

all X on MUVI disabled from trigger
all threshold of MUVI1 at 1 MeV to avoid as much as possible noise

LN filling problem solved !!!
pressure of the target slowly increasing about 820 mbar
We are filling the target to reach 1000 mbar


test cats1

Stop at 2019-06-29 13:03:26.602226
Stop comment: radioactive beam recovered
Fixede786s Sat Jun 29 13:08:17 2019 OttoRunRUNGeneralCryo-3HeRun81
*************************
Run : 81
*************************

Start at 2019-06-29 13:08:04.442155

target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted

AGATA position +51-25 mm

TRIGGER :
- CMG
- M2
- VAMOS%1000
- CATS%1000

all X on MUVI disabled from trigger
all threshold of MUVI1 at 1 MeV to avoid as much as possible noise

LN filling problem solved !!!
pressure of the target slowly increasing about 820 mbar
We are filling the target to reach 1000 mbar


test cats1

Stop at 2019-06-29 19:22:32.016536
Stop comment:
Fixede786s Sat Jun 29 20:31:21 2019 OttoRunRUNGeneralCryo-3HeRun84
*************************
Run : 84
*************************

Start at 2019-06-29 20:30:36.534403

target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted

AGATA position +51-25 mm

TRIGGER :
- CMG
- M2
- VAMOS%1000
- CATS%1000

all X on MUVI disabled from trigger
all threshold of MUVI1 at 1 MeV to avoid as much as possible noise

LN filling problem solved !!!
pressure of the target slowly increasing about 820 mbar
We are filling the target to reach 1000 mbar


test cats1

time calibration run
Fixede786s Mon Jul 1 11:00:16 2019 OttoRunRUNGeneralCryo-3HeRun85
*************************
Run : 85
*************************

Start at 2019-07-01 10:59:27.358229

target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted

AGATA position +51-25 mm

TRIGGER :
- CMG
- M2
- VAMOS%1000
- CATS%1000

all X on MUVI disabled from trigger
all threshold of MUVI1 at 1 MeV to avoid as much as possible noise

LN filling problem solved !!!
pressure of the target slowly increasing about 820 mbar
We are filling the target to reach 1000 mbar


test cats1

time calibration run

New line added just for clarity!!!

Bye Bye!
Salut!

Ok! Ok!

Stop at 2019-07-01 13:14:06.453759
Stop comment: type your message here

Stop at 2019-07-01 13:14:06.826283
Stop comment: type your message here

Stop at 2019-07-01 13:14:07.927090
Stop comment: type your message here

Stop at 2019-07-01 13:14:08.499190
Stop comment: type your message here

Stop at 2019-07-01 13:14:08.824522
Stop comment: type your message here

Stop at 2019-07-01 13:20:31.953407
Stop comment: type your message here

Stop at 2019-07-01 13:52:26.358087
Stop comment: