MUST2 SAMURAI_2018 S3-LEB-LPC SUPERNEMO MUGAST EXPAND SCALP GALATRON HiCARI VELO
  MUGAST, Page 13 of 127  Not logged in MUGAST
Fixed, e793s, Fri Mar 5 17:01:27 2021, mugast, S1, Hardware, N/A, Giving up on S1 
We mounted the new S1 detector from Surrey with a new custom cable on several MUFEE cards using the CsI ASICs.
We managed to put voltage on it and get decent leakage current. 
We modified the DAS parameters from CsI values to Si values (gain, shaping time, etc).
Fixed, TEST, Fri Mar 5 16:54:25 2021, OttoH.V., OTHER, Hardware, N/A, HVM_SUMMARY 
[SIZE=5][COLOR=green][B]*** hvm watcher summary ***[/B] [/COLOR][/SIZE] 
 [SIZE=2][COLOR=green]            [/COLOR] Slot: [COLOR=blue]8  [/COLOR] (A1821H) Channel:[COLOR=blue]1 [/COLOR]  Vmon: [COLOR=blue]    0 V  [/COLOR] Imon:
[COLOR=blue]    0 uA[/COLOR] Status: [COLOR=pink]Off[/COLOR]
Fixed, TEST, Fri Mar 5 16:54:24 2021, OttoH.V., OTHER, Hardware, N/A, HVM_SUMMARY 
[SIZE=5][COLOR=green][B]*** hvm watcher summary ***[/B] [/COLOR][/SIZE] 
 [SIZE=2][COLOR=green]+10V/t1234  [/COLOR] Slot: [COLOR=blue]3  [/COLOR] (A1516B) Channel:[COLOR=blue]5 [/COLOR]  Vmon: [COLOR=blue] 9.99 V  [/COLOR] Imon:
[COLOR=blue]0.159 A [/COLOR] Status: [COLOR=green]On[/COLOR]
Fixed, TEST, Fri Mar 5 16:54:24 2021, OttoH.V., OTHER, Hardware, N/A, HVM_SUMMARY 
[SIZE=5][COLOR=green][B]*** hvm watcher summary ***[/B] [/COLOR][/SIZE] 
 [SIZE=2][COLOR=green]+10v/mg     [/COLOR] Slot: [COLOR=blue]5  [/COLOR] (A1516B) Channel:[COLOR=blue]5 [/COLOR]  Vmon: [COLOR=blue]   10 V  [/COLOR] Imon:
[COLOR=blue]0.407 A [/COLOR] Status: [COLOR=green]On[/COLOR]
Fixed, e793s, Fri Mar 5 16:50:55 2021, OttoTime, RUN, General, N/A, TimeCalibratorRun 
*** Run Time Calibrator FSC ***
  - Run: 622
  - MUVI: MUVI1
Fixed, e793s, Fri Mar 5 16:39:08 2021, OttoRun, RUN, General, 3-alpha, Run620 
*************************
  [SIZE=3][B]Run  : 620[/B][/SIZE]
*************************
Fixed, e793s, Fri Mar 5 16:31:08 2021, OttoRun, RUN, General, 3-alpha, Run619 
*************************
  [SIZE=3][B]Run  :619[/B][/SIZE]
*************************
Fixed, e793s, Fri Mar 5 16:16:28 2021, mugast, DAQ, Software, , Pedestals for E793S 12x
Pedestal alignment was performed for Energy qnd Time for all cards with cooling at 0°. 
The values are attached here and saved in : cd ganacq_manip/mugast/das-save/pedestal-save/pedestal_5march2021_15h54/
Fixed, e793s, Fri Mar 5 16:14:21 2021, OttoRun, RUN, General, 3-alpha, Run618 MGRAWEY.pngMGRAWEX.pngMMRAWEY.pngMMRAWEX.png
*************************
  [SIZE=3][B]Run  : 618[/B][/SIZE]
*************************
Fixed, e775s, Thu Mar 4 10:45:42 2021, OttoRun, RUN, General, N/A, Run200 
*************************
  [SIZE=3][B]Run  : 200[/B][/SIZE]
*************************
Fixed, e793s, Wed Mar 3 17:18:42 2021, Franco, GRIT, General, N/A, MUVI's scheme and connections on the chamber connections_chamber_MUGAST.pngMUVI.PNG
 
Ongoing, e793s, Wed Mar 3 17:12:57 2021, Franco, S1, Hardware, N/A, Schema of the new S1 S1.PNG
The signals of the new S1 will be splitted among 6 Mugast boards, according to the scheme attached.
The HV for the S1 have been updated accordingly, with labels like MG11/CsIMGXX.
Fixed, e793s, Fri Feb 12 18:19:52 2021, OttoTime, RUN, General, N/A, TimeCalibratorRun 
*** Run Time Calibrator FSC ***
  - Run: 615
  - MUVI: MUVI2
Fixed, e793s, Fri Feb 12 18:12:00 2021, Adrien, S1, Hardware, N/A, Kapton change 
Arnaud open today and change the Y Kapton (YA5) of the S1. Everything under vacuum/cooled again so I ran FSC, hopefully this will fix the issue we have
with the Junction side.
Fixed, e793s, Fri Feb 12 18:07:24 2021, OttoTime, RUN, General, N/A, TimeCalibratorRun 
*** Run Time Calibrator FSC ***
  - Run: 614
  - MUVI: MUVI2
Fixed, e793s, Fri Feb 12 08:52:30 2021, mugast, RUN, General, 3-alpha, run 613: long alpha run  
all Muvi in trigger, mugast stand alone. all detector Bias on.
Fixed, e793s, Mon Feb 8 16:33:22 2021, OttoTime, RUN, General, N/A, TimeCalibratorRun 
*** Run Time Calibrator FSC ***
  - Run: 612
  - MUVI: MUVI2
Fixed, e793s, Mon Feb 8 16:20:42 2021, OttoTime, RUN, General, N/A, TimeCalibratorRun 
*** Run Time Calibrator FSC ***
  - Run: 611
  - MUVI: MUVI2
Fixed, e793s, Mon Feb 8 11:28:38 2021, OttoTime, RUN, General, N/A, TimeCalibratorRun 
*** Run Time Calibrator FSC ***
  - Run: 610
  - MUVI: MUVI3
Fixed, e793s, Mon Feb 8 11:25:29 2021, Adrien and Freddy, S1, General, N/A, FSC run fscE_08022023.pngfscT_08022022.png
we ran  FSC on all card. S1 is dismounted.
Pulser goes on all channel and works well so electronic is fine.
We will ask Laurent to remount the detector and re run the test.

ELOG Home Page