Fixed  Thu Feb 20 17:58:51 2020 FrancoGRITGeneral Time calibration for e775s exp
Attached the raw and calibrated time spectra for the pulser runs.
Attachment 1: T_raw_MM.png
T_raw_MM.png
Attachment 2: T_raw_MG.png
T_raw_MG.png
Attachment 3: T_cal_MM.png
T_cal_MM.png
Attachment 4: T_cal_MG.png
T_cal_MG.png
FixedTEST Sun Feb 23 20:50:06 2020 MarlèneGRITGeneral RUN 115
Screenshots from MUGAST

The impact matrix looks fine but there are sectors missing but in the raw we see 13 sectors (instead of 16)

There E-Tof spectra missing because TY is either not working but we checked that TX is working --> TX should be used for MG2, MG5 and MG6
The ToF seems to have 2 components : one when MUGAST has triggered and one when Vamos has triggered.
We see alpha contribution that come from fusino-evaporation on Au that can go up to 120 deg backward.

We put the BT and HT screenshots as a reference

The cooling is at 0 degree.... there is not much alcohol left in the cooler, we will refill tonight.
Attachment 1: e775s_impact_matrix.png
e775s_impact_matrix.png
Attachment 2: MG_TOF.png
MG_TOF.png
Attachment 3: MG_TX_raw.png
MG_TX_raw.png
Attachment 4: MG_TY_raw.png
MG_TY_raw.png
Attachment 5: BT_MG1.png
BT_MG1.png
Attachment 6: HT_MG.png
HT_MG.png
Attachment 7: MG_BT2.png
MG_BT2.png
Attachment 8: TAC.png
TAC.png
FixedTEST Sun Feb 23 22:17:49 2020 MarleneRUNHardware tracetek error
When filling the coolers with alcohol the tracetek cable triggered and switched off water and supplies.
Beam is off
FixedTEST Mon Feb 24 10:26:29 2020 MarlèneGRITGeneral infos
We have changed the range of the TAC FAG-HF to 500ns (it was 200 ns before)

there is no delay on the STOP of MUVI, it is directly the HF clean.

The inspection lines of MUVI1 are sent to vamos acquisition room.
FixedTEST Mon Feb 24 16:11:05 2020 Freddy, Franco, Didier, MarlèneRUNGeneral run 123 screenshots
RUN 123
CD2 300ug/cm2
Beam intensity 1e5 pps

reference run
Attachment 1: run123_theta_Elab.png
run123_theta_Elab.png
Attachment 2: run123_theta_Elab_unzoomed.png
run123_theta_Elab_unzoomed.png
Attachment 3: run123_TACs.png
run123_TACs.png
Attachment 4: run123_TofMG.png
run123_TofMG.png
Attachment 5: run123_impactMG.png
run123_impactMG.png
Fixed  Mon Feb 24 22:34:43 2020 Marlène NicolasRUNGeneral Screenshots run 126 (target CD2+Au #2 rotated 180 deg)
Attachment 1: run126_theta_Elab.png
run126_theta_Elab.png
Attachment 2: run126_Tof.png
run126_Tof.png
Fixede775s Tue Feb 25 11:50:16 2020 Marlène & FrancoOFFLINEGeneral PID with gate on T_FPMW_HF
The black points are what remains after putting the condition abs(T_FPMW_HF-13000)<1000, for MG1 and the annular.
Attachment 1: MG1_gatetime_run128.png
MG1_gatetime_run128.png
Attachment 2: annular_gatetime_run128.png
annular_gatetime_run128.png
Fixede775s Tue Feb 25 15:09:58 2020 mugastGRITGeneral Screenshots run 131
Attachment 1: run131_MG_impact.png
run131_MG_impact.png
Attachment 2: run131_MG_ETOF.png
run131_MG_ETOF.png
Attachment 3: run131_MG_TAC.png
run131_MG_TAC.png
Attachment 4: run131_MG_corr.png
run131_MG_corr.png
Attachment 5: run131_MG_STRX.png
run131_MG_STRX.png
Attachment 6: run131_MG_STRY.png
run131_MG_STRY.png
Fixed  Wed Feb 26 14:31:47 2020 Fairouz, Nicolas, MarlèneGRITGeneral RUN 137
Run 137
CD2+Au #2
Beam intensity 6e+5pps on average

Beam was retuned during the night and the quality of the beam is probably degraded
Fixede775s Wed Feb 26 15:02:38 2020 MarlèneRUNGeneral RUN 137
RUN 137
CD2+Au #2
beam intensity 6e+5pps

beam was retuned during the night, seems to be degraded although intensity is higher.
Attachment 1: Ex_137.pdf
Ex_137.pdf
Attachment 2: Ex_run128_137.pdf
Ex_run128_137.pdf
Fixed  Wed Feb 26 15:04:29 2020 mugastRUNGeneral beam retuning
Beam is retuned from PCP to improve its quality
Fixede775s Wed Feb 26 15:12:02 2020 MarlèneRUNGeneral RUN 137
Screenshots of RUN137


Marlène wrote:
RUN 137
CD2+Au #2
beam intensity 6e+5pps

beam was retuned during the night, seems to be degraded although intensity is higher.
Attachment 1: MG_ElabThetaLab_run137.png
MG_ElabThetaLab_run137.png
Attachment 2: MG_YT.png
MG_YT.png
Attachment 3: MG_XT.png
MG_XT.png
Attachment 4: MG_YE.png
MG_YE.png
Attachment 5: MG_XE_run137.png
MG_XE_run137.png
Attachment 6: Tac_run137.png
Tac_run137.png
Attachment 7: impact_MG_run137.png
impact_MG_run137.png
Attachment 8: ToF_MUGAST.png
ToF_MUGAST.png
Fixede775s Wed Feb 26 16:19:35 2020 MarlèneRUNGeneral RUN 137 (suite)
Here is the E gamma versus Excitation energy spectra for run 137



Marlène wrote:
RUN 137
CD2+Au #2
beam intensity 6e+5pps

beam was retuned during the night, seems to be degraded although intensity is higher.
Attachment 1: Ex_Egamma_run137.png
Ex_Egamma_run137.png
Fixede775s Wed Feb 26 20:41:21 2020 Nicolas, SylvainRUNGeneral run 138
Start at 15h45 (Feb. 26th)
Stop at 09h48 (Feb. 27th)

Beam: 19O
Intensity: 4e5 pps

Target; CD2 + Au #2
Fixede775s Wed Feb 26 20:45:14 2020 Nicolas, SylvainRUNGeneral run 138 - minutes of what happened during the run
Here is a brief summary of what happened over the course of the run:

From 4 pm to 12 pm (February 26th)
1. 20h00: call PCP because beam intensity decreased smoothly from 4e5 pps (16h) to 3e5 pps (20h) -> back to 3.5e5 pps
2. 20h25: call PCP because of sudden drop of intensity. VAMOS PF bias OK and beam OK according to PCP. Beam came back alone up to 3.2e5 pps with relaxation time of qbout 10 min as if some element was rumping up.
3. 23h50: FP_CATS tripped

From 12 pm to 8 am (February 27th)
Stable beam intensity at around 3.1e5 pps with VAMOS FP HV tripping 4 times

From 8 am to end of run (February 27th)
1. 08h05: FP_CATS0..5 tripped
2. 08h50: FP_CATS0..5 tripped


Nicolas, Sylvain wrote:
Start at 15h45
Stop at

Beam: 19O
Intensity: 4e5 pps

Target; CD2 + Au #2
Fixede775s Wed Feb 26 21:57:05 2020 Nicolas, SylvainRUNGeneral run 138 screenshots1

Nicolas, Sylvain wrote:
Start at 15h45
Stop at

Beam: 19O
Intensity: 4e5 pps

Target; CD2 + Au #2
Attachment 1: RUN138_scalers.png
RUN138_scalers.png
Attachment 2: RUN138_ThetaLab_ELab.png
RUN138_ThetaLab_ELab.png
Attachment 3: RUN138_impact.png
RUN138_impact.png
Attachment 4: RUN138_TAC.png
RUN138_TAC.png
Attachment 5: RUN138_ToF.png
RUN138_ToF.png
Fixede775s Thu Feb 27 03:13:33 2020 Faïrouz, Chloé & MichalRUNGeneral run 138 screenshots1 at 03:13

Faïrouz, Chloé, Michal wrote:

Faïrouz, Chloé, Michal wrote:
Screen shots at 03:15 am

Beam: 19O
Intensity: 3.4e5 pps

Target; CD2 + Au #2
Attachment 1: Scalers-run138-2020-02-27-03-08-33.png
Scalers-run138-2020-02-27-03-08-33.png
Attachment 2: MG-impact-matrix-2020-02-27_03-10-04.png
MG-impact-matrix-2020-02-27_03-10-04.png
Attachment 3: MG_Theta_E-2020-02-27_03-10-38.png
MG_Theta_E-2020-02-27_03-10-38.png
Attachment 4: TACS-2020-02-27_03-11-13.png
TACS-2020-02-27_03-11-13.png
Attachment 5: MG-E-TOF-2020-02-27_03-12-30.png
MG-E-TOF-2020-02-27_03-12-30.png
Attachment 6: Run-138-Ex-2020-02-27_03-17-51.png
Run-138-Ex-2020-02-27_03-17-51.png
Fixede775s Thu Feb 27 03:21:03 2020 Faïrouz, Chloé & MichalRUNGeneral run 138 screenshots at 07:20 am

Faïrouz, Chloé & Michal wrote:

Faïrouz, Chloé, Michal wrote:

Faïrouz, Chloé, Michal wrote:
Screen shots at 07:20 am

Beam: 19O
Intensity: 3.0e5 pps

Target; CD2 + Au #2
Attachment 1: Scaler-run138-2020-02-27_07-16-33.png
Scaler-run138-2020-02-27_07-16-33.png
Attachment 2: TAC-2020-02-27_07-17-16.png
TAC-2020-02-27_07-17-16.png
Attachment 3: MG-E-TOF-2020-02-27_07-17-43.png
MG-E-TOF-2020-02-27_07-17-43.png
Attachment 4: Impact-2020-02-27_07-18-17.png
Impact-2020-02-27_07-18-17.png
Attachment 5: MG-E-Theta-2020-02-27_07-18-50.png
MG-E-Theta-2020-02-27_07-18-50.png
Attachment 6: FPMW-HF-2020-02-27_07-19-39.png
FPMW-HF-2020-02-27_07-19-39.png
Attachment 7: Ex-run138-2020-02-27_08-19-27.png
Ex-run138-2020-02-27_08-19-27.png
Fixede775s Thu Feb 27 08:21:19 2020 Nicolas, SylvainRUNGeneral run 138 - Beam intensity time evolution

Nicolas, Sylvain wrote:
Start at 15h45
Stop at

Beam: 19O
Intensity: 4e5 pps

Target; CD2 + Au #2
Attachment 1: BeamIntensityProfile.png
BeamIntensityProfile.png
Attachment 2: BeamIntensityProfileNight.png
BeamIntensityProfileNight.png
Fixede775s Thu Feb 27 08:38:05 2020 Nicolas, SylvainOFFLINEGeneral run 138 - analysis

Nicolas, Sylvain wrote:
Start at 15h45
Stop at

Beam: 19O
Intensity: 4e5 pps

Target; CD2 + Au #2
Attachment 1: Ex-run138-2020-02-27_08-19-27.png
Ex-run138-2020-02-27_08-19-27.png
Attachment 2: RUN138_Ex.png
RUN138_Ex.png
Attachment 3: RUN138_ExEg.png
RUN138_ExEg.png

ELOG Home Page