Status |
ExpNbr |
Date |
Author |
Type |
Category |
Target-Source |
Subject |
Text |
 |
Pinned | e786s |
Wed Apr 10 13:30:48 2019 |
Valerian, Adrien and Antoine (edit by Louis) | OFFLINE | Software | N/A | How to perform analysis for e786s | [U][B]
Here is the updated procedure to convert
a run, with VAMOS data and add those data |
|
Fixed | |
Mon Jun 24 14:14:17 2019 |
Valerian, Adrien - Dudouet | OFFLINE | General | | p,p' analysis of run49 | |
  |
Fixed | e744s |
Mon Apr 15 12:15:22 2019 |
Valerian and Sylvain | OFFLINE | Software | CH2 | Kinematic line 2p | |
 |
Fixed | e744s |
Thu Apr 11 16:07:59 2019 |
Valerian and Louis | OFFLINE | Software | CH2 | HF resolution | Using the value of the HF=83,4ns. We can obtain
the calibrated plot of T_CATS2_HF which is
joined and gives us roughly a 2-3 ns resolution |
|
Fixed | TEST |
Wed Apr 3 15:49:02 2019 |
Valerian and Jenn, Chloé and François | OTHER | Hardware | N/A | Electronic scheme of VAMOS and MUGAST | Edit : 4/04/19 Valerian and Jenn small correction
of MUGAST scheme |
    |
Fixed | TEST |
Mon Mar 4 12:28:26 2019 |
Valerian and Iulian | MUST2 | General | 3-alpha | No DECS and acq error run 47-48-49. Das Saved after problem fixed | Fixed. Solution:
1) CENTRUM not configured on GENTRUM Master
2) GMT matrix on GMT3 (2nd slave) was all |
|
Fixed | e744s |
Sun Apr 14 12:11:52 2019 |
Valerian and Iulian | OFFLINE | Software | CH2 | VERY VERY preliminary results Gamma-p | Around 8-10 events are found in the good energy
vs Brho region and Si_T vs Si_E region for
the gamma emission.
|
  |
Fixed | e744s |
Thu Apr 11 18:20:27 2019 |
Valerian and Diego | OFFLINE | Software | CH2 | Timing issue between vamos and mugast | A cut is seen on the TAC between MUGAST and
the focal plane multiwire of VAMOS when one
select XF or DC0_Y>-1500
|
 |
Fixed | e744s |
Fri Apr 12 22:44:43 2019 |
Valerian and Diego | OFFLINE | Software | CH2 | 13N measured in Vamos do correspond to protons events in MUST2 | Doing a cut on M vs M/Q corresponding to 13N
one can see that most of the events do correspond
to protons events in MUST2!!! |
|
Fixed | e744s |
Wed Apr 17 13:48:25 2019 |
Valerian and Diego | RUN | Software | CH2 | New Dalitz plot with good angles in vamos | The previous Dalitz plot was wrong as I was
using the angle from vamos as polar angles.
Now that it is corrected we see that this |
|
Fixed | TEST |
Wed Mar 27 11:06:33 2019 |
Valerian and Cyril | GRIT | Hardware | 3-alpha | Evolution of resolution with temperature | For the mugast telescope, changing the temperature
from 13deg to minus 4deg doesn't seem to
have an impact as strong as for the MUST2 |
 |
Fixed | TEST |
Thu Mar 21 13:28:06 2019 |
Valerian and Adrien | DAQ | Software | N/A | First fully functionnal time calibration run on X and Y of MG1 | Our program is able to perform the time calibration
of our detector for both X and Y, see picture
attached. |
 |
Fixed | e786s |
Mon Jun 24 16:50:48 2019 |
Valerian and Adrien | OFFLINE | Software | Cryo-3He | 3He-d kinematics | To plot those kinematics, there is a macros
called kine.cxx
|
|
Fixed | e786s |
Wed Jun 26 09:56:17 2019 |
Valerian and Adrien | RUN | General | Cryo-3He | Identification E-TOF MUGAST | Light particle ID in MUGAST (no S1 because
shitty time).
|
 |
Fixed | e775s |
Wed Mar 4 11:06:49 2020 |
Valerian Irene | OFFLINE | General | N/A | Second 2+ | Current comparison between the CD2 target
and CD2 + Au target. |
|
Fixed | e744s |
Wed Apr 17 07:04:24 2019 |
Valerian (Mariluz was the one in charge of AGATA) | AGATA | Software | CH2 | Cristal 4A stopped | Cristal 4A was stopped during the night, see
AGATA logbook for more details. |
|
Fixed | e768s |
Wed Jul 31 14:32:53 2019 |
Valerian & Co | RUN | General | N/A | Beam intensity dropped | unstable beam since 15min... Intensity dropped
from 8e6 to 1e5
|
 |
Fixed | e786s |
Tue Jun 25 08:23:58 2019 |
Valerian | BEAM | General | N/A | Beam status | Currently there is no beam since 22h-> Thunder
+ no more 48Ca sample. The sample was out
during the night.
|
|
Fixed | TEST |
Wed Mar 6 15:02:51 2019 |
Valerian | MUST2 | Hardware | 3-alpha | Cooling temperature is relevant for the resolution | Using run 52 -> 13deg and run 87 -> -5deg.
We see that the resolution improves at -5 |
 |
Fixed | TEST |
Thu Mar 28 15:39:01 2019 |
Valerian | OFFLINE | Software | N/A | Status of the converion with VAMOS program | The conversion with VAMOS program was validated
using run_0020 from com2019
|
|