Pinnede786s Wed Apr 10 13:30:48 2019 Valerian, Adrien and Antoine (edit by Louis)OFFLINESoftwareN/AHow to perform analysis for e786s[U][B]
Here is the updated procedure to convert
a run, with VAMOS data and add those data
  
Fixed  Mon Jun 24 14:14:17 2019 Valerian, Adrien - DudouetOFFLINEGeneral p,p' analysis of run49   Screenshot_from_2019-06-24_14-09-39.pngKine46Arpp.pngGamma46Arpp.png 
Fixede744s Mon Apr 15 12:15:22 2019 Valerian and SylvainOFFLINESoftwareCH2Kinematic line 2p   ELabThetaLab_new.pngEcm2p_new.png 
Fixede744s Thu Apr 11 16:07:59 2019 Valerian and LouisOFFLINESoftwareCH2HF resolutionUsing the value of the HF=83,4ns. We can obtain
the calibrated plot of T_CATS2_HF which is
joined and gives us roughly a 2-3 ns resolution
 T_CATS2_HFns.png 
FixedTEST Wed Apr 3 15:49:02 2019 Valerian and Jenn, Chloé and François OTHERHardwareN/AElectronic scheme of VAMOS and MUGASTEdit : 4/04/19 Valerian and Jenn small correction
of MUGAST scheme
 VAMOS_electronic_scheme.pdfVAMOS_electronic_scheme.pptxMUGAST_Electronic_scheme_v2.pptxMUGAST_Electronic_scheme_v3.pdfMUGAST_Electronic_scheme_v3.pptx 
FixedTEST Mon Mar 4 12:28:26 2019 Valerian and IulianMUST2General3-alphaNo DECS and acq error run 47-48-49. Das Saved after problem fixedFixed. Solution:
1) CENTRUM not configured on GENTRUM Master
2) GMT matrix on GMT3 (2nd slave) was all
 run49_error_acq.png 
Fixede744s Sun Apr 14 12:11:52 2019 Valerian and IulianOFFLINESoftwareCH2VERY VERY preliminary results Gamma-pAround 8-10 events are found in the good energy
vs Brho region and Si_T vs Si_E region for
the gamma emission.
 EPCM_Brho_gamma.pngPossiblePafterGamma_ToF.pngEpcm_gamma.png 
Fixede744s Thu Apr 11 18:20:27 2019 Valerian and DiegoOFFLINESoftwareCH2Timing issue between vamos and mugastA cut is seen on the TAC between MUGAST and
the focal plane multiwire of VAMOS when one
select XF or DC0_Y>-1500
 T_MUGAST_FPMW.pngT_MUGAST_FPMW1.png 
Fixede744s Fri Apr 12 22:44:43 2019 Valerian and DiegoOFFLINESoftwareCH213N measured in Vamos do correspond to protons events in MUST2Doing a cut on M vs M/Q corresponding to 13N
one can see that most of the events do correspond
to protons events in MUST2!!!
 Vamos13N_and_MUST2protons.png 
Fixede744s Wed Apr 17 13:48:25 2019 Valerian and DiegoRUNSoftwareCH2New Dalitz plot with good angles in vamosThe previous Dalitz plot was wrong as I was
using the angle from vamos as polar angles.
Now that it is corrected we see that this
 DalitzGoodAngle.png 
FixedTEST Wed Mar 27 11:06:33 2019 Valerian and CyrilGRITHardware3-alphaEvolution of resolution with temperatureFor the mugast telescope, changing the temperature
from 13deg to minus 4deg doesn't seem to
have an impact as strong as for the MUST2
 MG4_13_deg.pngMG4_minus_4_deg.png 
FixedTEST Thu Mar 21 13:28:06 2019 Valerian and AdrienDAQSoftwareN/AFirst fully functionnal time calibration run on X and Y of MG1Our program is able to perform the time calibration
of our detector for both X and Y, see picture
attached.
 CALIMEROY1.pngFIRST_TIMECALIBRATION_WITHOUT_CALIMERO.png 
Fixede786s Mon Jun 24 16:50:48 2019 Valerian and AdrienOFFLINESoftwareCryo-3He3He-d kinematicsTo plot those kinematics, there is a macros
called kine.cxx
 Kine47K.png 
Fixede786s Wed Jun 26 09:56:17 2019 Valerian and AdrienRUNGeneralCryo-3HeIdentification E-TOF MUGASTLight particle ID in MUGAST (no S1  because
shitty time).
 ETOF_d.pdfETOF_d_vamos.pdf 
Fixede775s Wed Mar 4 11:06:49 2020 Valerian IreneOFFLINEGeneralN/ASecond 2+ Current comparison between the CD2 target
and CD2 + Au target.
 Second2PlusCurrentStat.pdf 
Fixede744s Wed Apr 17 07:04:24 2019 Valerian (Mariluz was the one in charge of AGATA)AGATASoftwareCH2Cristal 4A stoppedCristal 4A was stopped during the night, see
AGATA logbook for more details.
  
Fixede768s Wed Jul 31 14:32:53 2019 Valerian & CoRUNGeneralN/ABeam intensity droppedunstable beam since 15min... Intensity dropped
from 8e6 to 1e5
 index.jpgscalers_run160_11.png 
Fixede786s Tue Jun 25 08:23:58 2019 Valerian BEAMGeneralN/ABeam statusCurrently there is no beam since 22h-> Thunder
+ no more 48Ca sample. The sample was out
during the night.
  
FixedTEST Wed Mar 6 15:02:51 2019 ValerianMUST2Hardware3-alphaCooling temperature is relevant for the resolutionUsing run 52 -> 13deg and run 87 -> -5deg.

We see that the resolution improves at -5
 MM2X_13deg.pngMM2X_minus5deg.png 
FixedTEST Thu Mar 28 15:39:01 2019 ValerianOFFLINESoftwareN/AStatus of the converion with VAMOS programThe conversion with VAMOS program was validated
using run_0020 from com2019
  