Fixed, com2019, Wed Apr 3 16:24:37 2019, Adrien, CATS, Software, N/A, Calib file from T.Roger   
|
Here are the calibration file provided by T. Roger. |
Pinned, TEST, Tue Feb 26 09:30:21 2019, Freddy, OTHER, Hardware, N/A, Cabling - MUVI / HT-BT / SA-SL
|
Cabling scheme - to update until final |
Fixed, e744s, Thu Apr 11 00:57:34 2019, iulian, OTHER, Hardware, CH2, CH2 target thickness
|
target CH2 104 um (2nd position)
Brho 14N6+ CATS2 |
Fixed, e768s, Mon Jul 22 05:30:37 2019, Faïrouz & Sylvain, RUN, General, LiF, CF put because AGATA went crazy
|
CF put because AGATA went crazy.
The beam profile is shifted in comparison to the initial profile at midnight. The beam is touching somewhere.
We asked the beam operators to try to center the beam. |
Fixed, e744s, Tue Apr 16 08:23:38 2019, Marion, Dieter, CATS, Hardware, CH2, CATS tripping - lowering bias
|
8:24 Faraday in.
-5V on CATS.
550V.
|
Fixed, e786s, Fri Jun 28 08:57:21 2019, Freddy, CATS, General, Cryo-3He, CATS scaler not compatible with beam rate
|
The CATS scaler si now around 100-120 kHz while the beam rate is still 40 kHz.
I checked the CATS signal and there is clearly noise who induce double logic signals. Not all of them are validated by FAG.
|
Fixed, e744s, Fri Apr 12 08:19:31 2019, Marlène, OFFLINE, General, , CATS profiles RUN87    
|
|
Fixed, com2019, Fri Apr 5 14:54:09 2019, mugast, CATS, Hardware, N/A, CATS high voltage lowered by 20V
|
CATS high voltage lowered by 20V from 580V to 560V (on request of J. Pancin) |
Fixed, e744s, Thu Apr 11 19:47:15 2019, Marlène, OFFLINE, General, , CATS charges in RUN76 and RUN77    
|
Offline analysis to reconstruct the charges in X and Y for CATS.
The charges are lower in RUN77 because bias was reduced to 580V.
Wires seem to be all OK. |
Fixed, e793s, Thu Mar 11 04:34:58 2021, ablaihan, louis, CATS, Hardware, CD2, CATS and VAMOS tripped
|
at 4h15 CATS and VAMOS tripped probably due to too high beam intensity (~1.2x10^6 pps). We switched them on again at 4h30. All the 6 CATS in Board2 in HV
by 525 V instead of 530 V.
|
Fixed, e744s, Thu Apr 11 09:05:37 2019, mugast, CATS, General, N/A, CATS HV changed to 580
|
CATS HV changed from 600V to 580V. |
Fixed, e744s, Tue Apr 16 11:40:04 2019, mugast, CATS, Hardware, N/A, CATS HV
|
CATS2 HV was decreased to 540 V at 11h30 to avoid the overcurrent.
Need to monitor Efficiency, Time resolution as soon as the run restart with new target. |
Fixed, e786s, Mon Jun 24 03:08:49 2019, Marlène, RUN, General, Cryo-3He, Brho issue
|
We observe a change in the Brho of VAMOS : the charged states are divided in 2 !!!
Édit: it is now understood as ice water cracking on the upstream of the target. |
Fixed, e768s, Fri Jul 19 00:03:25 2019, Jenn, BEAM, General, , Brho information 15 Oxygen beam 4.7MeV/u
|
|
Fixed, e768s, Fri Jul 19 00:06:35 2019, Jenn, BEAM, General, , Brho information 15 Oxygen beam 4.7MeV/u
|
|
Fixed, e744s, Fri Apr 12 01:56:35 2019, Jenn, Louis, Cyril,Daniele and Chloe, BEAM, General, Empty, Brho for the new 14O setting RUN86 
|
CATS2 on and target position 1: hole
Beam intensity in CATS: 4E4
VAMOS intensity 1.3E4
|
Fixed, , Fri Apr 5 19:05:56 2019, Marlène Nicolas, GRIT, General, , Brho beam & 17O
|
Brho of 16O @ 6 MeV/A after 1mg/cm2 = 0.68906
Brho of 17O for g.s. after 1 mg/cm2=0.71065
Brho of 17O @ 871 keV after 1mg/cm2=0.70831 |
Fixed, e744s, Fri Apr 12 06:51:59 2019, Louis, BEAM, General, N/A, Brho and energy of the 14O beam
|
|
Fixed, e744s, Mon Apr 15 10:07:23 2019, Antoine, VAMOS, Hardware, , Brho VAMOS
|
During the change, Johan entered in G1 and connected a mass cable for the drift chambers. |
Fixed, e768s, Tue Jul 16 15:13:01 2019, Jenn, DIego, VAMOS, General, , Brho
|
|
|