MUST2 SAMURAI_2018 S3-LEB-LPC SUPERNEMO MUGAST EXPAND SCALP GALATRON HiCARI VELO
  MUGAST, Page 123 of 132  Not logged in MUGAST
Fixed, e793s, Tue Mar 9 10:47:12 2021, mugast, DAQ, Hardware, N/A, GMT config 
GMT

1-MUVI1
2-MUVI2
3-MUVI3
4-
5-CATS2clean % (/10**4 for the moment)
6-VAMOS% (/10 for the moment)
7-MUST2% (/10 for the moment)
8-
9-
10-VAMOS
11-OR AGATA
Fixed, e793s, Tue Mar 9 11:47:14 2021, Franco, RUN, General, N/A, TACS calibration TACS_calib.pngTACS_calib_fits.png
Coefficients uploaded in MUGAST/e793s/calibTACS_e793s.cal (set offsets to 0 and kept only the slope)
Fixed, e793s, Wed Mar 10 13:03:00 2021, OttoRun, RUN, General, 3-alpha, Run41 
*************************
Run : 41
*************************

Start at 2021-03-10 12:58:41.805693

run 3 alpha

Stop at 2021-03-10 13:06:49.699879
Stop comment:
Fixed, e793s, Wed Mar 10 13:09:12 2021, OttoRun, RUN, General, 3-alpha, Run42 
*************************
Run : 42
*************************

Start at 2021-03-10 13:09:04.739262

run 3 alpha

Stop at 2021-03-10 13:30:31.317348
Stop comment: changing trigger
Fixed, e793s, Wed Mar 10 13:33:31 2021, OttoRun, RUN, General, N/A, Run43 
*************************
Run : 43
*************************

Start at 2021-03-10 13:33:01.346673

Agata or as trigger for testing before beam

Stop at 2021-03-10 13:55:37.322216
Stop comment:
Fixed, e793s, Wed Mar 10 13:57:54 2021, OttoRun, RUN, General, N/A, Run45 
*************************
Run : 45
*************************

Start at 2021-03-10 13:57:50.701600

Agata or as trigger for testing before beam

Stop at 2021-03-10 14:01:53.157232
Stop comment:
Fixed, e793s, Wed Mar 10 14:43:14 2021, OttoRun, RUN, General, N/A, Run46 
*************************
Run : 46
*************************

Start at 2021-03-10 14:42:50.365714

Test run for Professor Clement. Beam in the room.

Stop at 2021-03-10 16:02:30.981195
Stop comment:
Fixed, e793s, Wed Mar 10 15:15:44 2021, Charlie, OTHER, General, , 48Ti Kinematics (additional) PDBarnes_Table_1.png48Ti_and_48K_Kinematics.pngWilton_mock-up.png
Supplementary to Adrien's kinematic calculations of 47Ti(d,p)48Ti, we have found a paper suggesting that by far the most strongly populated state in 48Ti is at 5.65 MeV. There is also a doublet of strong states at around 4.4 MeV, and two more strongly populated states at 6.06 and 6.33 MeV.

See attachment for the full table from P. D. Barnes (1965), Phys.Rev.138.B597.

Some quick CatKin models also show that a (theoretical) state at 7 MeV in 48Ti would almost perfectly overlap with the ground state of 48K. As such, when looking at the lab energy with the g.s. of 48K normalised to 0, we could expect to see these strongly populated 48Ti states at -0.67, -0.94, -1.35 and -2.6 MeV (for 6.3, 6.06, 5.65 and 4.4 MeV, respectively).

See the attached plot for kinematics of 47K(d,p)48K @ 0 MeV and 4 MeV, as well as 47Ti(d,p)48Ti @ 0 MeV, 5.65 MeV and the theoretical 7 MeV. Additionally, see a mock-up of the expected spectra from Wilton.
Fixed, e793s, Wed Mar 10 16:07:43 2021, OttoRun, RUN, General, N/A, Run47 
*************************
Run : 47
*************************

Start at 2021-03-10 16:07:22.660368

Beam in the room; restarting daq after recompiling the action

Stop at 2021-03-10 16:09:55.182936
Stop comment: stopping run to remove vamos% from trigger
Fixed, e793s, Wed Mar 10 16:12:40 2021, OttoRun, RUN, General, Empty, Run48 
*************************
Run : 48
*************************

Start at 2021-03-10 16:11:56.052849

Beam in the room; restarting daq after changing trigger.

Trigger are MUGAST/MM/CATS%

Stop at 2021-03-10 16:30:03.651340
Stop comment: stopping to remove cats% from trigger so we can set the timing
Fixed, e793s, Wed Mar 10 16:31:50 2021, OttoRun, RUN, General, CD2, Run49 
*************************
Run : 49
*************************

Start at 2021-03-10 16:31:31.923710

Beam in the room; restarting daq after changing trigger.
target: CD2 0.5mg
Trigger are MUGAST/MM

Stop at 2021-03-10 16:53:12.237279
Stop comment: stopping to change GMT windows
Fixed, e793s, Wed Mar 10 17:00:06 2021, OttoRun, RUN, General, CD2, Run50 
*************************
Run : 50
*************************

Start at 2021-03-10 16:59:44.368649

Beam in the room; restarting daq after changing trigger.
target: CD2 0.5mg
Trigger are MUGAST/MM
changing width of the FAG to 1us

Stop at 2021-03-10 17:12:21.612357
Stop comment: stopping to get in and change the cats delay
Fixed, e793s, Wed Mar 10 17:25:22 2021, OttoRun, RUN, General, CD2, Run51 
*************************
Run : 51
*************************

Start at 2021-03-10 17:24:45.316220

Beam in the room; restarting daq after changing trigger.
target: CD2 0.5mg
Trigger are MUGAST/MM
changed width of the FAG to 400ns
Removed extra 100ns delay from cats validation

Stop at 2021-03-10 18:14:26.996869
Stop comment: stop for VAMOS data analysis. Coherence check

Stop at 2021-03-10 20:46:36.348405
Stop comment: stopped the run to add cats% in the trigger
Fixed, e793s, Wed Mar 10 18:22:01 2021, Taras, Nishu, Chloe, RUN, General, , status screenshot 
Fixed, e793s, Wed Mar 10 19:23:37 2021, mugast, OFFLINE, General, CD2, Kinematics and Ex - RUN051 RUN51_ELabThethaLab_Ex.pngRUN51_ELabThethaLab_Ex_Egamma.png
Preliminary analysis results, Backward angles.

MM5 position are incorrect
Fixed, e793s, Wed Mar 10 20:53:55 2021, OttoRun, RUN, General, CD2, Run52 
*************************
Run : 52
*************************

Start at 2021-03-10 18:16:20.382279

Beam in the room; restarting daq after changing trigger.
target: CD2 0.5mg
Trigger are MUGAST/MM
changed width of the FAG to 400ns
Removed extra 100ns delay from cats validation
Fixed, e793s, Wed Mar 10 21:02:03 2021, OttoRun, RUN, General, CD2, Run53 
*************************
Run : 53
*************************

Start at 2021-03-10 21:00:45.001418

Production run.

Beam: 4.5e5 pps
target: CD2 0.5mg
Trigger are MUGAST/MM/CATS%

Stop at 2021-03-11 09:40:34.970341
Stop comment: run stopped to enter and check position 0 ATC6 trip
Fixed, e793s, Wed Mar 10 22:19:22 2021, Adrien, Freddy, Wilton and Nigel, OFFLINE, General, CD2, Offline analysis of run 51 and 52 using ganil files RUN5152_ELabThethaLab.png
We noted that the gamma ray efficiency calculated from file converted from AGATA was close 100%. We concluded that the conversion program was reauiring a gamma ray to be present.

This is confirm by analysis of the ganil dta file that show 10 time more count that the agata one.

Here is a kinematical plot of the run 51 and 52 analysised from ganil file.

There is a 1.5MeV state present that disappear in the agata coincidence. This is to be investigated with more data. Is it an isomer state? from 48K? 48Ti?
Fixed, e793s, Thu Mar 11 01:24:47 2021, Ablaihan, Louis, RUN, General, CD2, run 53 screenshots 16x
Screenshots of the run 53
Beam intensity is ~ 7.4 10^5 pps in CATS and 5.3 10^5 in VAMOS
~75% transmition
Dt is ~ 3%
Fixed, e793s, Thu Mar 11 02:50:14 2021, Ablaihan, Louis, BEAM, General, CD2, VAMOS transmition  
at 2h50 beam intensity reached 1.1x10^6 pps in CATS for 6.5x10^5 pps in VAMOS. The transmition decreased to 60% without any shift of the beam position. Does it come from a loss of efficiency of VAMOS at high counting rate?

ELOG Home Page