  MUGAST, Page 123 of 132  Not logged in MUGAST
Status ExpNbr Date Author Typedown Category Target-Source Subject Text Attachments
FixedTEST Wed Jan 22 15:53:25 2020 MarlèneGRITHardware Changes for MUGAST2020Changes were made for MUGAST 2020 : 
1- the board for MG10 before (2nd square)
is now for MG6 new trap
  
Fixed  Wed Jan 22 18:39:54 2020 Marlène & FrancoGRITHardware Tests of the new boxThe new box has been tested together with
the new CAEN SY4527 power supply. We connected
a laptop with Linux to the power supply by
 box_pin_map.pdf 
FixedTEST Thu Jan 23 14:06:51 2020 Franco&MarlèneGRITHardware3-alphaStatus of detectors on 23-01-20Short alpha source run (not saved on disk)
to test the detectors.
 6x 
FixedTEST Mon Feb 10 18:42:23 2020 Franco&MarlèneGRITHardware New MUGAST detectors MG2 MG6MG2 and MG6 have been mounted in the chamber.

MG2 reference number is 3227_10 thickness=468um
  
Fixed  Fri Feb 14 10:26:46 2020 Franco&MarlèneGRITHardware HT BT screenshots before e775sDetectors are cooled down to 0 deg.  BT_CAEN2_MGnew.pngBT_CAEN1_part1_MM.pngBT_CAEN1_part2_MG.pngHV_part1_MM_MG.pngHV_part2_MG.png 
FixedTEST Tue Feb 18 17:53:36 2020 FrancoGRITGeneral3-alphaAlpha calibration for exp e775s (run 0501)In attachment raw and calibrated spectra for
run 0501 (alpha source) for MUST2 and Mugast.
 E_raw_MM.pngE_raw_MG.pngE_cal_MM.pngE_cal_MG.png 
Fixed  Thu Feb 20 17:58:51 2020 FrancoGRITGeneral Time calibration for e775s expAttached the raw and calibrated time spectra
for the pulser runs. 
 T_raw_MM.pngT_raw_MG.pngT_cal_MM.pngT_cal_MG.png 
FixedTEST Sun Feb 23 20:50:06 2020 MarlèneGRITGeneral RUN 115 Screenshots from MUGAST

The impact matrix looks fine but there are
 8x 
FixedTEST Mon Feb 24 10:26:29 2020 MarlèneGRITGeneral infosWe have changed the range of the TAC FAG-HF
to 500ns (it was 200 ns before)
  
Fixede775s Tue Feb 25 15:09:58 2020 mugastGRITGeneral Screenshots run 131   6x 
Fixed  Wed Feb 26 14:31:47 2020 Fairouz, Nicolas, MarlèneGRITGeneral RUN 137 Run 137
CD2+Au #2
Beam intensity 6e+5pps on average
  
Fixede775s Thu Feb 27 12:35:39 2020 Marlène NicolasGRITGeneral RUN128-137_138 analysisKinematic lines for the full statistics up
to now
States above E*>7 MeV are cut in the angles
 Elab_Thetalab_RUN128_137_138.pngExEg_RUN128_137_139.pngEx_RUN128_137_138.png 
Fixed  Fri Feb 28 11:20:22 2020 MarlèneGRITGeneral RUN 150RUN 150
 target CD2 300 ug/cm2
Beam intensity 4e+5pps
  
Fixede775s Sat Feb 29 14:32:39 2020 mugast : ValerianGRITHardwareN/AScheme of the Mugast position in the nptool referentialThe detectors with a cross show the problematic
detectors
 EEDD6F02-7FF9-41B0-BB0B-673611AF018D.jpeg 
FixedTEST Tue Mar 10 11:05:40 2020 FrancoGRITHardware3-alphaTrip in the cooling system after e775s experimentIn the weekend after the experiment e775s
(7-8 March 2020) a problem with the cooling
system of Mugast occurred. In particular,
 run0193_12_heat.pngrun0193_13_heat_MG6vsMG1.pngrun_0524_MG11XYraw.pngrun_0193_9_19_11_MG11XYraw.png 
Fixede793s Tue Mar 10 18:46:13 2020 Franco & MarlèneGRITHardware Changes in MUGAST- MG6 (FZ) is disconnected.
- MG7 (500um) is dismounted and MG8 (375um)
is replacing it.
 MG_X_raw_11032020.pngMG_Y_raw_11032020.png 
FixedTEST Wed Mar 11 18:12:40 2020 Marlène & FrancoGRITHardwareN/AUpdate on Mugast(Probably) final update of the status of the
detector before the experiment:
  
Fixede793s Wed Mar 11 19:02:54 2020 mugastGRITGeneral3-alphaAlpha calibration for e793s - run 600========================================
========================================
========================================
  
Fixede793s Wed Mar 11 22:01:18 2020 mugastGRITGeneral3-alphaAlpha calibration - run 600After 2h30' of run we checked the temperatures
of the boards and we noticed that MG7 had
63 degrees on both X and Y, which is very
 MG7X_run600_Eevolution.png 
FixedTEST Thu Mar 12 10:14:02 2020 Marlène & FrancoGRITHardwareN/ADebugging the cooling systemWe realized that the increase of temperature
of MG7 and MG9 during run 600 (alpha calibration)
was due to the fact that a loop of the cooling
  