Status |
ExpNbr |
Date |
Author |
Type |
Category |
Target-Source |
Subject |
Fixed | e793s |
Tue Mar 9 10:47:12 2021 |
mugast | DAQ | Hardware | N/A | GMT config |
GMT
1-MUVI1
2-MUVI2
3-MUVI3
4-
5-CATS2clean % (/10**4 for the moment)
6-VAMOS% (/10 for the moment)
7-MUST2% (/10 for the moment)
8-
9-
10-VAMOS
11-OR AGATA |
Fixed | e793s |
Tue Mar 9 11:47:14 2021 |
Franco | RUN | General | N/A | TACS calibration |
Coefficients uploaded in MUGAST/e793s/calibTACS_e793s.cal (set offsets to 0 and kept only the slope) |
Fixed | e793s |
Wed Mar 10 13:03:00 2021 |
OttoRun | RUN | General | 3-alpha | Run41 |
*************************
Run : 41
*************************
Start at 2021-03-10 12:58:41.805693
run 3 alpha
Stop at 2021-03-10 13:06:49.699879
Stop comment: |
Fixed | e793s |
Wed Mar 10 13:09:12 2021 |
OttoRun | RUN | General | 3-alpha | Run42 |
*************************
Run : 42
*************************
Start at 2021-03-10 13:09:04.739262
run 3 alpha
Stop at 2021-03-10 13:30:31.317348
Stop comment: changing trigger |
Fixed | e793s |
Wed Mar 10 13:33:31 2021 |
OttoRun | RUN | General | N/A | Run43 |
*************************
Run : 43
*************************
Start at 2021-03-10 13:33:01.346673
Agata or as trigger for testing before beam
Stop at 2021-03-10 13:55:37.322216
Stop comment: |
Fixed | e793s |
Wed Mar 10 13:57:54 2021 |
OttoRun | RUN | General | N/A | Run45 |
*************************
Run : 45
*************************
Start at 2021-03-10 13:57:50.701600
Agata or as trigger for testing before beam
Stop at 2021-03-10 14:01:53.157232
Stop comment: |
Fixed | e793s |
Wed Mar 10 14:43:14 2021 |
OttoRun | RUN | General | N/A | Run46 |
*************************
Run : 46
*************************
Start at 2021-03-10 14:42:50.365714
Test run for Professor Clement. Beam in the room.
Stop at 2021-03-10 16:02:30.981195
Stop comment: |
Fixed | e793s |
Wed Mar 10 15:15:44 2021 |
Charlie | OTHER | General | | 48Ti Kinematics (additional) |
Supplementary to Adrien's kinematic calculations of 47Ti(d,p)48Ti, we have found a paper suggesting that by far the most strongly populated state in 48Ti is at 5.65 MeV. There is also a doublet of strong states at around 4.4 MeV, and two more strongly populated states at 6.06 and 6.33 MeV.
See attachment for the full table from P. D. Barnes (1965), Phys.Rev.138.B597.
Some quick CatKin models also show that a (theoretical) state at 7 MeV in 48Ti would almost perfectly overlap with the ground state of 48K. As such, when looking at the lab energy with the g.s. of 48K normalised to 0, we could expect to see these strongly populated 48Ti states at -0.67, -0.94, -1.35 and -2.6 MeV (for 6.3, 6.06, 5.65 and 4.4 MeV, respectively).
See the attached plot for kinematics of 47K(d,p)48K @ 0 MeV and 4 MeV, as well as 47Ti(d,p)48Ti @ 0 MeV, 5.65 MeV and the theoretical 7 MeV. Additionally, see a mock-up of the expected spectra from Wilton. |
Fixed | e793s |
Wed Mar 10 16:07:43 2021 |
OttoRun | RUN | General | N/A | Run47 |
*************************
Run : 47
*************************
Start at 2021-03-10 16:07:22.660368
Beam in the room; restarting daq after recompiling the action
Stop at 2021-03-10 16:09:55.182936
Stop comment: stopping run to remove vamos% from trigger |
Fixed | e793s |
Wed Mar 10 16:12:40 2021 |
OttoRun | RUN | General | Empty | Run48 |
*************************
Run : 48
*************************
Start at 2021-03-10 16:11:56.052849
Beam in the room; restarting daq after changing trigger.
Trigger are MUGAST/MM/CATS%
Stop at 2021-03-10 16:30:03.651340
Stop comment: stopping to remove cats% from trigger so we can set the timing |
Fixed | e793s |
Wed Mar 10 16:31:50 2021 |
OttoRun | RUN | General | CD2 | Run49 |
*************************
Run : 49
*************************
Start at 2021-03-10 16:31:31.923710
Beam in the room; restarting daq after changing trigger.
target: CD2 0.5mg
Trigger are MUGAST/MM
Stop at 2021-03-10 16:53:12.237279
Stop comment: stopping to change GMT windows |
Fixed | e793s |
Wed Mar 10 17:00:06 2021 |
OttoRun | RUN | General | CD2 | Run50 |
*************************
Run : 50
*************************
Start at 2021-03-10 16:59:44.368649
Beam in the room; restarting daq after changing trigger.
target: CD2 0.5mg
Trigger are MUGAST/MM
changing width of the FAG to 1us
Stop at 2021-03-10 17:12:21.612357
Stop comment: stopping to get in and change the cats delay |
Fixed | e793s |
Wed Mar 10 17:25:22 2021 |
OttoRun | RUN | General | CD2 | Run51 |
*************************
Run : 51
*************************
Start at 2021-03-10 17:24:45.316220
Beam in the room; restarting daq after changing trigger.
target: CD2 0.5mg
Trigger are MUGAST/MM
changed width of the FAG to 400ns
Removed extra 100ns delay from cats validation
Stop at 2021-03-10 18:14:26.996869
Stop comment: stop for VAMOS data analysis. Coherence check
Stop at 2021-03-10 20:46:36.348405
Stop comment: stopped the run to add cats% in the trigger |
Fixed | e793s |
Wed Mar 10 18:22:01 2021 |
Taras, Nishu, Chloe | RUN | General | | status screenshot |
|
Fixed | e793s |
Wed Mar 10 19:23:37 2021 |
mugast | OFFLINE | General | CD2 | Kinematics and Ex - RUN051 |
Preliminary analysis results, Backward angles.
MM5 position are incorrect |
Fixed | e793s |
Wed Mar 10 20:53:55 2021 |
OttoRun | RUN | General | CD2 | Run52 |
*************************
Run : 52
*************************
Start at 2021-03-10 18:16:20.382279
Beam in the room; restarting daq after changing trigger.
target: CD2 0.5mg
Trigger are MUGAST/MM
changed width of the FAG to 400ns
Removed extra 100ns delay from cats validation |
Fixed | e793s |
Wed Mar 10 21:02:03 2021 |
OttoRun | RUN | General | CD2 | Run53 |
*************************
Run : 53
*************************
Start at 2021-03-10 21:00:45.001418
Production run.
Beam: 4.5e5 pps
target: CD2 0.5mg
Trigger are MUGAST/MM/CATS%
Stop at 2021-03-11 09:40:34.970341
Stop comment: run stopped to enter and check position 0 ATC6 trip |
Fixed | e793s |
Wed Mar 10 22:19:22 2021 |
Adrien, Freddy, Wilton and Nigel | OFFLINE | General | CD2 | Offline analysis of run 51 and 52 using ganil files |
We noted that the gamma ray efficiency calculated from file converted from AGATA was close 100%. We concluded that the conversion program was reauiring a gamma ray to be present.
This is confirm by analysis of the ganil dta file that show 10 time more count that the agata one.
Here is a kinematical plot of the run 51 and 52 analysised from ganil file.
There is a 1.5MeV state present that disappear in the agata coincidence. This is to be investigated with more data. Is it an isomer state? from 48K? 48Ti? |
Fixed | e793s |
Thu Mar 11 01:24:47 2021 |
Ablaihan, Louis | RUN | General | CD2 | run 53 screenshots |
Screenshots of the run 53
Beam intensity is ~ 7.4 10^5 pps in CATS and 5.3 10^5 in VAMOS
~75% transmition
Dt is ~ 3% |
Fixed | e793s |
Thu Mar 11 02:50:14 2021 |
Ablaihan, Louis | BEAM | General | CD2 | VAMOS transmition |
at 2h50 beam intensity reached 1.1x10^6 pps in CATS for 6.5x10^5 pps in VAMOS. The transmition decreased to 60% without any shift of the beam position. Does it come from a loss of efficiency of VAMOS at high counting rate? |