Fixed, , Tue Mar 12 10:21:17 2019, Marlène, MUST2, General, , Debug Calimero Y test_GECO_masquage_mate2&10activees_seulement.png
Debug of Calimero

test of channels (mate) masking : OK with DAS, OK with GECCO (see attachment) when done in the "disable channels" of GECCO
Fixed, com2019, Tue Apr 2 17:15:47 2019, Freddy, OTHER, General, N/A, DWBA Cross section 16Odp17O_s12_6AMeV.dat
16O(d,p)17O*(1/2+) @ 6 MeV/nucleon - SF = 1
Fixed, e793s, Fri Mar 12 20:43:38 2021, Cyril, Louis, OFFLINE, General, CD2, DWBA DWBA.pngDWBA2.png
DBWA analysis have been performed with TWOFNR, here the comparison with data for the 0, 2 and 4 MeV peaks. 
1p1/2 (black line) and 0f7/2 (red line)
Fixed, , Sun Jul 21 23:18:25 2019, Kseniia, AGATA, General, , Current total stats on 19Ne  58.png
Runs 112, 115, 116, 117, 118, 119, 120 and ~1h of 121

We see 3 peak candidates with energies not matching anything known
Fixed, e768s, Sun Jul 21 17:45:54 2019, kseniia, AGATA, General, , Current total stats on 19Ne run112-119part-c.pngrun112-119part_zoom.png
Stats from runs 112, 115, 116, 117, 118, 119(~15h)
beta = 8%
Q = 9 & 10
Fixed, , Mon Jul 22 19:13:45 2019, Kseniia, AGATA, General, , Current stats on 19Ne 112-125.png
Runs 112..125, still no joy...
Fixed, TEST, Thu Feb 21 16:43:45 2019, Valérian, Freddy, Iulian and Marlène, GRIT, Hardware, N/A, Current measurements Depletion_Mugast_MG.pdfDepletion_Mugast_M2.pdfDepletion_Mugast.xlsx
Joined are the the I(V).

Detectors seems to respond accordingly with the previous measurements from Marlène. 
Fixed, e768s, Sat Jul 27 16:54:29 2019, mugast, AGATA, General, LiF, Current full stat on 19Ne  Runs112-154part.pngRuns112to154part_zoom.png
Identified in VAMOS, all the good runs from 112 up to partial 154
Nothing new, a few counts near 2497
Fixed, com2019, Tue Apr 9 10:44:40 2019, Valerian, MUST2, Software, CD2, CsI Calib files from the commissioning for proton Cal_CsI_E_MM1.calCal_CsI_E_MM2.calCal_CsI_E_MM3.calCal_CsI_E_MM4.cal
CsI calibration obtained from the proton data of the commissioning of mugast.
Fixed, e744s, Mon Apr 15 01:49:49 2019, Freddy, MUST2, General, N/A, CsI - low energy suppression run125_2_MUGAST_3_4.png
The suppression zone around the pedestals for the CsI was set to 8210 since the beginning of the experiment.

It corresponds to about 30 channels above the zero (8192). The gain for the CsI is of 40 keV/channel.
Fixed, e744s, Wed Apr 17 07:04:24 2019, Valerian (Mariluz was the one in charge of AGATA), AGATA, Software, CH2, Cristal 4A stopped 
Cristal 4A was stopped during the night, see AGATA logbook for more details.
Fixed, e768s, Fri Jul 19 13:09:14 2019, mugast, BEAM, General, LiF, Counting rates as a function of beam intensity IMG_20190719_130411.jpg
checking counting rates as a function of increasing beam intensity
Fixed, e744s, Sun Apr 14 14:23:15 2019, Nikola Dieter Nicolas Beyhan Kseniia, ONLINE, General, CH2, Count rates Fag.pngMugast.pngAgata.pngVAMOS.pngCats1.png
 
Fixed, TEST, Wed Mar 6 15:02:51 2019, Valerian, MUST2, Hardware, 3-alpha, Cooling temperature is relevant for the resolution MM2X_13deg.pngMM2X_minus5deg.png
Using run 52 -> 13deg and run 87 -> -5deg.

We see that the resolution improves at -5 degree. See attached picture as an example for MM2 X.
Pinned, com2019, Fri Apr 5 23:34:46 2019, Adrien et Valerian, OFFLINE, General, N/A, Converting data and adding stat to Kine Plot 
Here is a procedure to convert a run and add data the Kinematic plot.

Every 4Go there will be a file change, convert the file and add it to the chain using the following step:
Fixed, e744s, Fri Apr 12 14:13:11 2019, Giuseppe D'Agata, OTHER, Software, N/A, Converted runs 

Runs additionally converted up to now:
70
Fixed, e768s, Sun Jul 21 16:42:20 2019, Marlène, OFFLINE, General, , Conversion of data from AGATA :update 
When VAMOS VXI crashed (run 118) the Action file was recompiled.
As AGATA conversion is not copying the ACTION file for each run, it used the old version by default (from 16th of july).
Ongoing, e775s, Sat Feb 29 03:26:52 2020, mugast, RUN, General, N/A, Continues RUN 153 RUN153_online0305.pngRUN153_online0306.pngRUN153_online0307.pngRUN153_online0308.pngRUN153_online0309.png
Run 153 (target #2 CD2+Au)

Beam Intensity 4*10⁵
Fixed, e744s, Thu Apr 11 11:36:09 2019, Valerian, OFFLINE, Software, CH2, Comparison between the two e744 current_exp_old_e744.png
Current calibration is off, a shift of -1.5MeV was done to reproduce the previous data.
Fixed, e744s, Fri Apr 12 10:12:45 2019, Valerian, OFFLINE, Software, CH2, Comparison between HF and CATS for ELab ThetaLab ELabThetaLabrun87_E744_HFvsCATS.png
Using the HF show noises that is not seen with CATS on the ToF spectra