Status ExpNbr Date Author Type Category Target-Sourcedown Subject
Fixede793s Fri Feb 12 08:52:30 2021 mugastRUNGeneral3-alpharun 613: long alpha run
all Muvi in trigger, mugast stand alone. all detector Bias on.
Fixede793s Fri Mar 5 16:14:21 2021 OttoRunRUNGeneral3-alphaRun618
*************************
Run : 618
*************************

Start at 2021-03-05 16:13:37.198228

run 3 alpha with pedestal and mugast DAQ
Attachment 1: MGRAWEY.png
MGRAWEY.png
Attachment 2: MGRAWEX.png
MGRAWEX.png
Attachment 3: MMRAWEY.png
MMRAWEY.png
Attachment 4: MMRAWEX.png
MMRAWEX.png
Fixede793s Fri Mar 5 16:31:08 2021 OttoRunRUNGeneral3-alphaRun619
*************************
Run :619
*************************

Start at 2021-03-05 16:30:39.722907

run 3 alpha without pedestals for MG but with pedestals for MM
Fixede793s Fri Mar 5 16:39:08 2021 OttoRunRUNGeneral3-alphaRun620
*************************
Run : 620
*************************

Start at 2021-03-05 16:38:16.476202

run 3 alpha without pedestal for all detectors, to be used for experiment

Stop at 2021-03-05 16:56:13.657279
Stop comment: type your message here

Stop at 2021-03-05 16:56:14.690583
Stop comment: type your message here

Stop at 2021-03-05 16:56:15.050212
Stop comment: type your message here

Stop at 2021-03-05 16:56:16.082040
Stop comment:

Stop at 2021-03-05 17:13:59.790265
Stop comment:

Stop at 2021-03-05 17:15:58.879697
Stop comment:

Stop at 2021-03-06 13:25:23.248954
Stop comment:

Stop at 2021-03-06 13:25:23.504734
Stop comment:

Stop at 2021-03-06 13:25:23.873972
Stop comment:
Fixede793s Sat Mar 6 13:25:38 2021 OttoRunRUNGeneral3-alphaRun630
*************************
Run : 630
*************************

Start at 2021-03-06 13:15:07.038908

run 3 alpha. time calibrator is on but is not giving any trigger so just sending random stop

Stop at 2021-03-08 09:50:00.232468
Stop comment:

Stop at 2021-03-08 15:28:10.213189
Stop comment:

Stop at 2021-03-08 15:28:10.881754
Stop comment:

Stop at 2021-03-08 15:28:11.447832
Stop comment:
Fixede793s Mon Mar 8 19:01:26 2021 FrancoGRITGeneral3-alphaEnergy and time calibrations before experiment
Energy resolutions:

MG1: 45 keV
MG2: 45 keV
MG3: 45 keV
MG4: 50 keV
MG5: 50 keV
MG7: 50 keV
Total Mugast: 50 keV

MM1: 50 keV
MM2: 50 keV
MM3: 45 keV
MM4: 50 keV
Total MUST2: 50 keV

MM5 (90 deg): 65 keV
Attachment 1: Mugast_Eraw_run630.png
Mugast_Eraw_run630.png
Attachment 2: MUST2_Eraw_run630.png
MUST2_Eraw_run630.png
Attachment 3: Mugast_Traw_fsc.png
Mugast_Traw_fsc.png
Attachment 4: MUST2_Traw_fsc.png
MUST2_Traw_fsc.png
Attachment 5: Mugast_Ecal_run630.png
Mugast_Ecal_run630.png
Attachment 6: MUST2_Ecal_run630.png
MUST2_Ecal_run630.png
Attachment 7: Mugast_Tcal_fsc.png
Mugast_Tcal_fsc.png
Attachment 8: MUST2_Tcal_fsc.png
MUST2_Tcal_fsc.png
Fixede793s Tue Mar 9 09:43:51 2021 OttoRunRUNGeneral3-alphaRun200
*************************
Run : 200
*************************

Start at 2021-03-08 17:43:53.831102

run 3 alpha

Stop at 2021-03-09 09:43:55.629095
Stop comment: type your message here

Stop at 2021-03-09 09:43:56.774034
Stop comment: type your message here

Stop at 2021-03-09 09:43:57.137511
Stop comment: type your message here

Stop at 2021-03-09 09:43:57.480913
Stop comment: type your message here
Fixede793s Wed Mar 10 13:03:00 2021 OttoRunRUNGeneral3-alphaRun41
*************************
Run : 41
*************************

Start at 2021-03-10 12:58:41.805693

run 3 alpha

Stop at 2021-03-10 13:06:49.699879
Stop comment:
Fixede793s Wed Mar 10 13:09:12 2021 OttoRunRUNGeneral3-alphaRun42
*************************
Run : 42
*************************

Start at 2021-03-10 13:09:04.739262

run 3 alpha

Stop at 2021-03-10 13:30:31.317348
Stop comment: changing trigger
Fixede793s Fri Mar 19 09:54:58 2021 Adrien, Freddy and FrancoGRITGeneral3-alphaAlpha calibration and noise on MG3 and MG4 after the experiment
While starting the alpha source runs after the experiment we found unexpected noise on MG3 and MG4 which cannot be excluded just by masking some Y mates (The X mates were already masked for both of them). We cannot figure out the origin of this noise.

Run 652 and 653 -> alpha runs without pedestal suppression
Run 655 -> alpha run during the night with pedestal suppression. Here MG3 and MG4 are not in the trigger.
Run 658 -> alpha run after an attempt to mask some noisy strips. The number of ORD is anyway high (some 10^4) but we see some alphas on MG3 and MG4.
Run 660 -> same as 658 but with higher thresholds on X (8250) and Y (7600 on MG3 and 7950 on MG4) to write less on disk but actually nothing changes. Therefore we are writing about 2.7 GB per hour.

The other runs (654, 656, 657 and 659) can be discarded.
Fixede744s Wed Apr 17 16:53:53 2019 Sylvain (AGATA)AGATAGeneral152EuRun 137 : Eu calibration
Run stared 2019/04/17

EU efficiency Calibration
Fixede786s Tue Jun 25 10:51:59 2019 OttoRunRUNGeneral152EuRun57
*************************
Run : 57
*************************

Start at 2019-06-25 10:50:03.672363

152Eu source @ AGATA

Stop at 2019-06-25 14:36:40.442175
Stop comment: end ; source removed
Fixed  Wed Jan 23 12:24:10 2019 MarlèneGRITGeneral Change cable MG1T - Exchange of detectors named MG9C & MG10C
23/01/19

Change of the HT/BT cable of MG1T (the one that had a short cut before when the chamber was closed)

MG9C is now the square detector nb 4 (Micron number) and coupled with the thick detector --> this detector was working properly
MG10C is detector nb 3 (Micron number) --> this detector was noisy before
Fixed  Fri Jan 25 17:02:25 2019 MarlèneGRITGeneral Detectors numbering
Attachment 1: MUGAST_detectors.pdf
MUGAST_detectors.pdf
FixedTEST Mon Mar 4 16:24:51 2019 valerian marlèneGRITGeneral RUN52 3 alpha source 4MUST2 no pedestal
RUN52
3 alpha source run with only 4MUST2 (in MUVI3) with pedestal suppression at 13 deg.
Attachment 1: HV_MUST2_during_run52.png
HV_MUST2_during_run52.png
Attachment 2: HV_MUST2_afternightrun52.png
HV_MUST2_afternightrun52.png
Attachment 3: 3alpha_source_4MUST2_nightrun52.png
3alpha_source_4MUST2_nightrun52.png
Fixed  Mon Mar 4 17:22:29 2019 MarleneGRITGeneral Temperature on MUFEE MUST2 13deg
Temperature given by DAS on MUVI3 (MUST2)
screenshots before run
after run : same values
Attachment 1: temperature_MM1_cooling13deg.png
temperature_MM1_cooling13deg.png
Attachment 2: temperature_MM1_Y_cooling13deg.png
temperature_MM1_Y_cooling13deg.png
Attachment 3: temperature_MM2_X_cooling13deg.png
temperature_MM2_X_cooling13deg.png
Attachment 4: temperature_MM2_Y_cooling13deg.png
temperature_MM2_Y_cooling13deg.png
Attachment 5: temperature_MM3_X_cooling13deg.png
temperature_MM3_X_cooling13deg.png
Attachment 6: temperature_MM3_Y_cooling13deg.png
temperature_MM3_Y_cooling13deg.png
Attachment 7: temperature_MM4_X_cooling13deg.png
temperature_MM4_X_cooling13deg.png
Attachment 8: temperature_MM4_Y_cooling13deg.png
temperature_MM4_Y_cooling13deg.png
FixedTEST Thu Mar 7 09:29:47 2019 MarleneGRITGeneral run 96 double alpha source run shifted 60mm
RUN 96

double alpha source
MG1
MG3
MG4
MG5

MG7
-
-
MG11

MM1
MM2
MM3
MM4

Missing strips in Y the same as in run 40

Problem with the annular MG11 -> no current when bias is put... --> to be checked

Problem with the insertion of the target holder --> solved

I also checked again MG9 (still noisy)
and MG10 with 140V I=0,52uA and counting rate is the same order of magnitude as trapezoid. Spectra are OK (not saved).
Attachment 1: run96_MG7_MG11notworking.png
run96_MG7_MG11notworking.png
Attachment 2: run96_3alphasource_shifted_MG1-5.png
run96_3alphasource_shifted_MG1-5.png
Attachment 3: run96_HV2.png
run96_HV2.png
Attachment 4: run96_HV1.png
run96_HV1.png
Fixed  Tue Mar 12 10:21:17 2019 MarlèneMUST2General Debug Calimero Y
Debug of Calimero

test of channels (mate) masking : OK with DAS, OK with GECCO (see attachment) when done in the "disable channels" of GECCO

With calimero, when we are on Y, when we click right to mask a mate we see 'disable mate 2' instead of 'disable mate 10' !!!
Attachment 1: test_GECO_masquage_mate2&10activees_seulement.png
test_GECO_masquage_mate2&10activees_seulement.png
Fixed  Tue Mar 19 19:35:53 2019 marleneGRITGeneral RUN 98 3alpha source all detectors
RUN 98
3 alpha source
all detectors (trap,square, MUST2,annular)
Temp=15deg
Acquisition mugast
Fixed  Thu Apr 4 17:55:47 2019 MarlèneGRITGeneral Scalers
The scalers are

Scaler 1:
1-FAG
2-OR MUGAST
3-OR AGATA
4-OR VAMOS
5-
6-
7-100Hz pulser
8-DT n pulser 100Hz
9-OR BACK
10-OR MUST2
11-OR SQUARE
12-AGATA DIV

Scaler 2:
17-CATS1
18-CATS1 n FAG
19-CATS2