Status |
ExpNbr |
Date |
Author |
Type |
Category |
Target-Source |
Subject |
Fixed | e793s |
Tue Mar 16 08:39:46 2021 |
Sarah, Chloé, Ablaihan | BEAM | General | CD2 | Beam dropped due to electric defect. @ 08h40 back @ 8h45 |
|
Fixed | e793s |
Tue Mar 16 11:09:23 2021 |
Sarah, Chloé, Ablaihan | RUN | General | CD2 | run63 screenshots at 11h05 |
|
Fixed | e793s |
Tue Mar 16 14:15:55 2021 |
Sarah, Chloé, Ablaihan | RUN | General | CD2 | run63_13h58_Screenshots |
|
Fixed | e793s |
Tue Mar 16 16:59:48 2021 |
Wilton & Franco | RUN | General | CD2 | Screnshots run 63 at 17h |
|
Fixed | e793s |
Tue Mar 16 17:47:27 2021 |
Wilton & Franco | RUN | General | CD2 | run63 screenshots at 17h |
|
Fixed | e793s |
Tue Mar 16 19:57:42 2021 |
Wilton & Franco | RUN | General | CD2 | Screenshots run 63 at 19h30 |
|
Fixed | e793s |
Tue Mar 16 21:46:58 2021 |
Wilton & Franco | RUN | General | CD2 | run63 screenshots at 21h30 |
|
Fixed | e793s |
Tue Mar 16 23:45:43 2021 |
Wilton & Franco | RUN | General | CD2 | run63 screenshots at 23h30 |
|
Fixed | e793s |
Tue Mar 16 23:57:29 2021 |
Charlie | RUN | General | CD2 | run63 scaler graph at 00h00 |
The scaler graph had frozen at 15:45, so all of the screenshots of it since then show the same section. Refreshed graph of the last few hours attached. |
Fixed | e793s |
Wed Mar 17 01:51:16 2021 |
Laetitia, Charlie | RUN | General | CD2 | run63 screenshots at 01h35 |
|
Fixed | e793s |
Wed Mar 17 03:48:24 2021 |
Laetitia, Charlie | RUN | General | CD2 | run63 screenshots at 03h45 (17/03) |
|
Fixed | e793s |
Wed Mar 17 05:49:02 2021 |
Laetitia, Charlie | RUN | General | CD2 | run 63 screenshots at 05h45 |
|
Fixed | e793s |
Wed Mar 17 07:18:39 2021 |
Laetitia, Charlie | ONLINE | General | CD2 | VAMOS acquisition stopped |
The acquisition on Vamos get stucked at 6h49. At 07:18, we stopped and started a new run (run 64) |
Fixed | e793s |
Wed Mar 17 07:21:23 2021 |
OttoRun | RUN | General | CD2 | Run64 |
*************************
Run : 64
*************************
Start at 2021-03-17 07:20:12.195520
run64. ACQ DT
Restart due to VAMOS data file not writing
Beam: 4.0e5 pps
target: CD2 0.5mg
Trigger are MUGAST/MM/CATS%
Stop at 2021-03-17 10:05:42.524724
Stop comment: End of e793s |
Fixed | e793s |
Wed Mar 17 07:25:24 2021 |
Laetitia, Charlie | ONLINE | General | CD2 | VAMOS acquisition stopped |
Vamos acquisition get stucked at 6h49 (with 2 Gbit). We stopped and started a new run at 07h18 (Run #64) |
Fixed | e793s |
Wed Mar 17 07:52:37 2021 |
Laetitia, Charlie | RUN | General | CD2 | run 64 screenshots at 07h45 |
|
Pending | e793s |
Wed Mar 17 08:03:15 2021 |
Laetitia, Charlie | AGATA | Hardware | CD2 | Agata issue |
There is a problem on Agata (the dewars are not being filled), and the Ge's start to warm up. The beam has been stopped at ~08h05. |
Fixed | e793s |
Wed Mar 17 09:53:38 2021 |
Cyril | AGATA | Hardware | CD2 | Agata issue |
Agata problem has been solved, a valve was closed by mistake.
Beam returned at 9h40. |
Fixed | e793s |
Wed Mar 17 10:23:44 2021 |
Cyril | RUN | General | CD2 | run 64 screenshots at 10h05 - End of e793s |
The beam/acq has been stopped around 10h05. |
Fixed | e793s |
Wed Mar 17 10:54:14 2021 |
OttoTime | RUN | General | N/A | TimeCalibratorRun |
*** Run Time Calibrator FSC ***
- Run: 635
- MUVI: MUVI1
- Side : MUFEE X (Junction)
- Telescope: 1, 2, 3, 4 |