Status ExpNbr Date Authordown Type Category Target-Source Subject Text Attachments
Fixede744s Tue Apr 16 14:13:17 2019 FreddyRUNGeneralCH2Log and scalers RUN 134 @ 14h13   18x 
Fixede786s Mon Jun 24 13:57:46 2019 FreddyGRITHardwareCryo-3HeTiming change: -50 ns on CATS stopWe removed 50 ns of delay to the CATS signal
used for the stop of the TOF (CATS-MUGAST/MUST)
because we were at the edge of the FAG.
 TimingCATSMUGAST.jpg 
Fixede786s Mon Jun 24 20:39:54 2019 FreddyRUNGeneralCryo-3Herun 56 screenshots   scalers_run55_24062019_20h37.png 
Fixede786s Wed Jun 26 03:46:09 2019 FreddyRUNGeneralCryo-3HeRun 58 -screenshotsScreenshots at 3h30 on the 26th of june during
RUN 58
 6x 
Fixede786s Wed Jun 26 03:55:20 2019 FreddyOFFLINEGeneralN/ARun conversion with AGATATo convert a run including the AGATA data,
here is the procedure (Only if you know reasonably
what you are doing ;-)
  
Fixede786s Wed Jun 26 04:54:07 2019 FreddyAGATAGeneralN/AAGATA validation problem (solved)Since about 4h20, we realized that AGATA validated
events and timestamps have suddenly dropped
on the AGATA display (yellow bars instead
  
Fixede786s Wed Jun 26 05:58:23 2019 FreddyOFFLINEGeneralN/AOffline analysis run 56 and 58   run56-58_analysis.pngrun56-59_analysis_gammarays.pngrun56_analysis.pngrun58_analysis.png 
Fixede786s Wed Jun 26 07:01:00 2019 FreddyAGATAGeneralN/AAGATA validation problem - again - during run 5930 min after Emmanuel left (around 6h00),
the AGATA problem of validation reappeared...
  
Fixede786s Wed Jun 26 07:04:06 2019 FreddyCATSGeneralN/AUpdate of CATS calibrationsNew calibrations in :
/home/mugast/analysis/nptool/Projects/calibration/cats/CATS2_X.vamos2019_e786s.npt
/home/mugast/analysis/nptool/Projects/calibration/cats/CATS2_X.vamos2019_e786s.npt
 run56_cats_newcal.pngrun_58_cats_newcal.png 
Fixede786s Fri Jun 28 08:54:50 2019 FreddyRUNGeneralCryo-3Herun 74[quote="Daniele Mengoni"]screenshots run 74
attached
  
Fixede786s Fri Jun 28 08:57:21 2019 FreddyCATSGeneralCryo-3HeCATS scaler not compatible with beam rateThe CATS scaler si now around 100-120 kHz
while the beam rate is still 40 kHz.
 ScalerCATSproblem.png 
Fixede768s Thu Jul 25 17:19:52 2019 FreddyOFFLINEGeneralN/AScaler evolution + scaler file readerAttached is a plot of several scaler (frequency
as a function of time) for all data from
run 120 to 140. 
 ScalerDisplay_r120_140.png 
Fixede768s Sat Jul 27 04:50:45 2019 FreddyOFFLINEGeneralN/AScaler summary since r120run 120 - 152

/home/mugast/analysis/scalereader/out/Chain.C
 scaler_update.png 
Fixede768s Sat Jul 27 11:49:40 2019 FreddyS1HardwareN/AAnnular degradationI posted a detailed entry during the night
about the evolution of the annular detector
but it was deleted somehow...
 annular_diag_run139_0.pdfannular_diag_run140_0.pdfannular_diag_run149_0.pdfannular_diag_run152_0.pdfHV7.pdf 
Fixed  Mon Feb 24 08:46:15 2020 FreddyOFFLINEGeneralCD2r116 - check GATCONFAfter a quick offline check, it seems that
we get MUGAST data only if MUGAST has triggered
(GATCONFMASTER = 1,2,4).
 r116_GATCONFMASTER.pngr116_DSSDE_GATCONFMASTER.png 
Fixed  Fri Apr 12 02:05:34 2019 François, Chloé TARGETGeneral Target changeChanged from Target 1 hole to Target 2 (104microns).   
Fixede744s Fri Apr 12 04:13:49 2019 François, ChloéOTHERHardware TAC to do listTAC of 
-  MUGAST-HF : HFvalidated is suspected to
come every 200ns after MUGAST, it suggests
  
Fixede768s Thu Jul 25 03:58:19 2019 François et GilbertRUNGeneralLiFSpectra Run 135 25/07/19 4:00Second snapshot of run 135
Spectra refreshed
 6x 
Fixede768s Mon Jul 22 14:45:34 2019 François D, FranckBEAMGeneralLiFBeam profileIntensity = 1.3e7 pps

Beam position shift compared to this morning
 Beam_profile_run125_22July2019_14h39.jpg 
Fixede744s Mon Apr 8 15:17:28 2019 FrançoisTARGETGeneral E744 Targets
Targets mounted on holder
 20190408_143035.jpg 