Status |
ExpNbr |
Date |
Author |
Type |
Category |
Target-Source |
Subject |
Text |
 |
Fixed | e775s |
Fri Feb 28 22:58:31 2020 |
Franck, Sylvain | ONLINE | General | CD2 | MUGAST Spectra Run 153 | Run 153 (target #2 CD2+Au)
MUGAST Spectra reset at start of run (~ 20h) |
   |
Ongoing | e775s |
Sat Feb 29 07:35:21 2020 |
mugast : Valerian | AGATA | General | CD2 | AGATA is not responding | Since ~4:30am AGATA is not responding, everyhting
was frozen. Unclear if it comes from disk
problem or other... |
|
Ongoing | e775s |
Sat Feb 29 14:27:48 2020 |
mugast ; Valerian | OFFLINE | Software | CD2 | Issues in the EX reconstruction | [quote="mugast ; valerian irene"]We are currently
working on the EX reconstruction which is
off for detector 1, 3 and 7... One of our |
6x |
Pinned | e793s |
Wed Mar 10 13:10:14 2021 |
Adrien | BEAM | General | CD2 | Kinematical line from Ti contamination | In black the Kinematical line for states known
to be populated strongly in 47Ti(d,p).
In blue range of the states of interrest.
|
 |
Fixed | e793s |
Wed Mar 10 16:31:50 2021 |
OttoRun | RUN | General | CD2 | Run49 | *************************
[SIZE=3][B]Run : 49[/B][/SIZE]
*************************
|
|
Fixed | e793s |
Wed Mar 10 17:00:06 2021 |
OttoRun | RUN | General | CD2 | Run50 | *************************
[SIZE=3][B]Run : 50[/B][/SIZE]
*************************
|
|
Fixed | e793s |
Wed Mar 10 17:25:22 2021 |
OttoRun | RUN | General | CD2 | Run51 | *************************
[SIZE=3][B]Run : 51[/B][/SIZE]
*************************
|
|
Fixed | e793s |
Wed Mar 10 19:23:37 2021 |
mugast | OFFLINE | General | CD2 | Kinematics and Ex - RUN051 | Preliminary analysis results, Backward angles.
MM5 position are incorrect |
 |
Pending | e793s |
Wed Mar 10 20:36:19 2021 |
Nigel | BEAM | General | CD2 | Beam centering - MUST2 hit pattern |
We made a quick check of the beam centering
by inspecting the MUST2 (forward angles) |
|
Fixed | e793s |
Wed Mar 10 20:53:55 2021 |
OttoRun | RUN | General | CD2 | Run52 | *************************
[SIZE=3][B]Run : 52[/B][/SIZE]
*************************
|
|
Fixed | e793s |
Wed Mar 10 21:02:03 2021 |
OttoRun | RUN | General | CD2 | Run53 | *************************
[SIZE=3][B]Run : 53[/B][/SIZE]
*************************
|
|
Fixed | e793s |
Wed Mar 10 22:19:22 2021 |
Adrien, Freddy, Wilton and Nigel | OFFLINE | General | CD2 | Offline analysis of run 51 and 52 using ganil files | We noted that the gamma ray efficiency calculated
from file converted from AGATA was close
100%. We concluded that the conversion program |
|
Fixed | e793s |
Thu Mar 11 01:24:47 2021 |
Ablaihan, Louis | RUN | General | CD2 | run 53 screenshots | Screenshots of the run 53
Beam intensity is ~ 7.4 10^5 pps in CATS
and 5.3 10^5 in VAMOS
|
16x |
Fixed | e793s |
Thu Mar 11 02:50:14 2021 |
Ablaihan, Louis | BEAM | General | CD2 | VAMOS transmition | at 2h50 beam intensity reached 1.1x10^6 pps
in CATS for 6.5x10^5 pps in VAMOS. The transmition
decreased to 60% without any shift of the |
|
Fixed | e793s |
Thu Mar 11 03:37:45 2021 |
Louis | OFFLINE | General | CD2 | Offline analyis run 53 | 1) Some angular distribution were extracted
(without any efficiency corrections) with
gate on Ex btw 0 and 2 MeV (where p orbitals |
    |
Fixed | e793s |
Thu Mar 11 04:34:58 2021 |
ablaihan, louis | CATS | Hardware | CD2 | CATS and VAMOS tripped | at 4h15 CATS and VAMOS tripped probably due
to too high beam intensity (~1.2x10^6 pps).
We switched them on again at 4h30. All the |
|
Fixed | e793s |
Thu Mar 11 04:38:31 2021 |
Ablaihan, Louis | BEAM | General | CD2 | Beam stopped due to a equipment tripped in CSS1 | Beam stopped 4h30 due to a equipment tripped
in CSS1. They are going to fix it. Estimated
timing of the intervention ~10/20min.
|
|
Fixed | e793s |
Thu Mar 11 05:12:36 2021 |
Ablaihan, Louis | RUN | General | CD2 | run 53 new screenshots | |
  |
Pending | e793s |
Thu Mar 11 09:02:20 2021 |
mugast | BEAM | Hardware | CD2 | beam interruption | power supply trip on the accelerator beam
is off for about 30min |
|
Fixed | e793s |
Thu Mar 11 09:04:38 2021 |
adrien | ONLINE | General | CD2 | Elastic scattering in MM5 | looking at E vs strip number in MM5 we see
clearly the pp and dd elastic scattering. |
|