Fixede768s Sun Jul 21 11:49:59 2019 Christiqn qnd DidierRUNGeneralLiFRun 119 screenshots
Fixede768s Sun Jul 21 14:54:28 2019 MarlèneRUNGeneral run 119 scalers
Fixede768s Sun Jul 21 16:39:38 2019 Marlène, AntoineBEAMGeneral IMPORTANT :How to send scalers to PCP
Go to PCP

Open a terminal
> ssh -X mugast@ganp154
passwd : Spiral1+

> ./LaunchScalers.sh
Fixede768s Sun Jul 21 16:42:20 2019 MarlèneOFFLINEGeneral Conversion of data from AGATA :update
When VAMOS VXI crashed (run 118) the Action file was recompiled.
As AGATA conversion is not copying the ACTION file for each run, it used the old version by default (from 16th of july).

We have modified the do_global_replay.sh so that it takes the ACTION file from the Replay directory where we have copied the ACTION file for run_0119.

So for conversion of file before run 119, the ACTION file in the zworkingDir of AGATA should be used.
For all conversion of runs after run 119, the ACTION file in the Replay directory should be used.
Fixed  Sun Jul 21 16:51:30 2019 Marlène, Christian, KseniaBEAMGeneral trial of tracking of the implantation of 15O
Looking at AGATA count rate, we have seen that the highest counting rate for 511 keV line is seen in the inner ring (0*A clusters).
The outer ring detectors (10B for example) counts 10 times less.

Our conclusion is that the beam is partly implanted at the entrance of MUGAST chamber.
The outer ring probably sees the implantation on MUST2 masks.

A more refined analysis is needed to backtrack where the 15O are implanted.
Fixede768s Sun Jul 21 17:02:25 2019 MarlèneBEAMGeneral beam intensity in the afternoon
Fixede768s Sun Jul 21 17:45:54 2019 kseniiaAGATAGeneral Current total stats on 19Ne
Stats from runs 112, 115, 116, 117, 118, 119(~15h)
beta = 8%
Q = 9 & 10
Fixede768s Sun Jul 21 17:57:08 2019 ChristianOTHERGeneral Previous Papers - gamma decays from states in the 4-MeV region

In comment to previous entry (see thread), on the 2019 Hall paper on 19Ne transitions:

Note that Fig 4.c is described as: Gated on the 238-keV 5/2+ to ground-state transition. The 3897.5-keV transition from the 4141.8-keV state is observed for the first time. This means the spectrum is designed such that it wold not show the 4033 transition. Therefore, whether they are sensitive to this is an open question. See also PhD thesis of Hall for further details?
Fixede768s Sun Jul 21 20:13:57 2019 Kseniia, Franco & NicolasBEAMGeneralN/ABeam intensity check

Didier & Nicolas wrote:

Jenn & Franck wrote:
OR Vamos: 1300 /s
Vamos gas profiler: 1.7x10^7 pps


OR VAMOS: ~ 1280 Hz
VAMOS gas profiler: 1.7e7 pps


OR VAMOS: ~ 1700 Hz
VAMOS gas profiler: 6e6 pps

beam moved in the horizontal plane (compare last 2 pictures of beam profiler)
Fixede768s Sun Jul 21 21:22:58 2019 OttoRunRUNGeneralLiFRun120
*************************
Run : 120
*************************

Start at 2019-07-21 21:19:46.530036

beam intensity after optimization 1.6e7 pps
OR VAMOS ~ 1500 Hz

Target: 1.25 mg/cm2 (nb 1) 7LiF

Triggers
MUGAST (backward, MM, 90 deg)
VAMOS %100
BaF2 %100

Stop at 2019-07-21 21:52:56.667021
Stop comment:
Fixede768s Sun Jul 21 22:17:18 2019 OttoRunRUNGeneralLiFRun121
*************************
Run : 121
*************************

Start at 2019-07-21 21:54:19.075515

beam intensity after optimization 1.6e7 pps
OR VAMOS ~ 1500 Hz

Target: 1.25 mg/cm2 (nb 1) 7LiF

Triggers
MUGAST (backward, MM, 90 deg)
VAMOS %100
BaF2 %100

Stop at 2019-07-22 05:09:23.310656
Stop comment: Problems with AGATA. Need to stop acquisition to recover AGATA validation
Fixed  Sun Jul 21 22:28:11 2019 KseniiaAGATAGeneral 15O implantation
Attached is the intensity distribution of 511 keV in run 119 (first image) vs run 120 (second image).
In between the two runs, beam optics was twitched.

These are the segments of crystal 02A which is near the beam pipe. Bottom is the tip of the detector, top is the
tail.
In run 119 it looks that the pillars that are near the beam line (columns 4 and 5) were seeing 511 more or less
homogeneously, which means that we are implanting 15O all over the pipe.
In run 120 looks like we still implant in the pipe, but also we implant a lot more at the target level.
Fixede768s Sun Jul 21 22:40:23 2019 Kseniia & NicolasBEAMGeneralN/Abeam check run121
best intensity ever: 2.2e7 pps
Fixed  Sun Jul 21 23:18:25 2019 KseniiaAGATAGeneral Current total stats on 19Ne
Runs 112, 115, 116, 117, 118, 119, 120 and ~1h of 121

We see 3 peak candidates with energies not matching anything known
Fixede768s Mon Jul 22 00:13:08 2019 Faïrouz & SylvainRUNGeneralLiFScreenshots run 121 @00h13
Fixede768s Mon Jul 22 02:07:44 2019 Faïrouz & SylvainBEAMGeneralLiFRun 121 : Beam profile
Status at 2:05 AM
Fixede768s Mon Jul 22 04:15:53 2019 Faïrouz & SylvainRUNGeneralLiFBeam intensity & profile & scalers
Fixede768s Mon Jul 22 05:03:59 2019 Faïrouz & SylvainRUNGeneralLiFScreenshots run 121 @05h04
Fixede768s Mon Jul 22 05:12:13 2019 OttoRunRUNGeneralLiFRun122
*************************
Run : 122
*************************

Start at 2019-07-22 05:11:56.885037

beam intensity after optimization 2.0e7 pps
OR VAMOS ~ 1500 Hz

Target: 1.25 mg/cm2 (nb 1) 7LiF

Triggers
MUGAST (backward, MM, 90 deg)
VAMOS %100
BaF2 %100

Stop at 2019-07-22 07:01:32.414292
Stop comment: stop run 122. We wait for Bertrand to optimize the beam centering and size.
Fixede768s Mon Jul 22 05:16:22 2019 Faïrouz & SylvainRUNGeneralLiFStop run 121 because problems of AGATA trigger validation
The run 121 has to be stopped because problems of AGATA trigger validation due to very high request rate of AGATA >=85K