Here is a todo list for today:
Data conversion:
- Raw run files are located in /data/mugastX/mugast/acquisition/run/
- Converting FSC run: 622 to 629 : run_0630.dat.06-03-21_13h15m06s
- Converting alpha run: 630. warning: there is 3 files for this run :run_0630.dat.06-03-21_13h15m06s and run_0630.dat.06-03-21_13h15m06s.1 and run_0630.dat.06-03-21_13h15m06s.2
Extract histogram for calibration: each macro need to be run twice, once for MUGAST and once for MUST2 if I remember correctly.
- Extract histograms for FSC run with the macro located : /home/mugast/analysis/nptool/Projects/calibration/Time/ExtractRawHisto_T.C
- Extract histograms for alpha run with the macro located : /home/mugast/analysis/nptool/Projects/calibration/Energy/ExtractRawHisto_E.C
Analysed the extracted histograms to calibrate: change the appropriate field in the macro before running
- FSC: /home/mugast/analysis/nptool/Projects/calibration/Time/TimeCalibrator.C
- alpha: /home/mugast/analysis/nptool/Projects/calibration/Energy/EnergyCalibrator.C
Prepare the calibration file to be used for the online/offline analysis:
- List all the file you just generated in /home/mugast/analysis/nptool/Projects/MUGAST/Calibration.txt
- Do some cleanup in the file so it is more readable.
- Check for missing/noisy strips and adjust config file in /home/mugast/analysis/nptool/Projects/MUGAST/configs
Analysed the calibration run with the new calibration:
- check T:strip and E:strip
- Post a summary of the resolution for each detector
Time calibrator on all tac
- Check the range of each tac
- wired the start/stop to the TAC inputs
- do an acquisition with enough statitics
- analyse it and put the calibration file in Calibration.txt
|