MUST2
SAMURAI_2018
S3-LEB-LPC
SUPERNEMO
MUGAST
EXPAND
SCALP
GALATRON
HiCARI
VELO
|
MUGAST |
Not logged in |
|
|
Message ID: 5943
Entry time: Tue Mar 10 18:46:13 2020
|
Author: |
Franco & Marlène |
ExpNbr: |
e793s |
Type: |
GRIT |
Category: |
Hardware |
Status: |
Fixed |
Target-Source: |
|
Subject: |
Changes in MUGAST |
|
|
- MG6 (FZ) is disconnected.
- MG7 (500um) is dismounted and MG8 (375um) is replacing it.
- The Y kapton of the annular has been changed.
Attached the alpha source spectra acquired on March 11th after the changes described above.
- MG8 looks better than MG7 but it looks like a part on X that was missing on MG7 is also missing on MG8, it can be due to a kapton.
- The annular behaves exactly as before and during e775s, the change of kapton did not solve the problem.
- MG9 here was not biased (it is on the MUST2 box). |
|
|
|