MUST2 SAMURAI_2018 S3-LEB-LPC SUPERNEMO MUGAST EXPAND SCALP GALATRON HiCARI VELO
  MUGAST  Not logged in MUGAST
Message ID: 5378     Entry time: Sat Jul 27 11:49:40 2019
Author: Freddy 
ExpNbr: e768s 
Type: S1 
Category: Hardware 
Status: Fixed 
Target-Source: N/A 
Subject: Annular degradation 
I posted a detailed entry during the night about the evolution of the annular detector but it was deleted somehow...

Signals on 3-4 segments were entirely lost between run 139 and 140. It creates a lot of holes on the impact matrix.

An increase of leakage current from 0.2 to 1.4uA occurred at about the same moment:

*Wed Jul 24 21:44:44 2019 --> 0.2 uA
*Thu Jul 25 06:53:15 2019 --> 0.6 uA
*Thu Jul 25 13:23:48 2019 --> 1.4 uA
*Thu Jul 25 22:25:35 2019 --> 1.3 uA
*Fri Jul 26 07:26:23 2019 --> 1.9 uA

Screenshots about this issues are in the e768s screenshot folder.

Edit : Adding some related files from the folder, not sure if the good ones (Sylvain)
Attachment 1: annular_diag_run139_0.pdf  184 kB  Uploaded Sun Jul 28 08:55:35 2019  | Hide | Hide all | Show all
annular_diag_run139_0.pdf
Attachment 2: annular_diag_run140_0.pdf  372 kB  Uploaded Sun Jul 28 08:55:43 2019  | Show | Hide all | Show all
Attachment 3: annular_diag_run149_0.pdf  397 kB  Uploaded Sun Jul 28 08:55:53 2019  | Hide | Hide all | Show all
annular_diag_run149_0.pdf
Attachment 4: annular_diag_run152_0.pdf  360 kB  Uploaded Sun Jul 28 08:56:03 2019  | Hide | Hide all | Show all
annular_diag_run152_0.pdf
Attachment 5: HV7.pdf  960 kB  Uploaded Mon Jul 29 10:58:45 2019  | Hide | Hide all | Show all
HV7.pdf

ELOG Home Page