By looking in details to the impact matrix of the annular detector, one can see that more segments than before are missing.
There was a clear degradation between run135 and run140.
HVmonitor summary of leakage current for MG11A:
*Wed Jul 24 21:44:44 2019 ---> 0.2 uA
*Thu Jul 25 06:53:15 2019 ---> 0.6 uA (run135 stop)
*Thu Jul 25 13:23:48 2019 ---> 1.4 uA (during run 140)
*Thu Jul 25 22:25:35 2019 ---> 1.3 uA
*Fri Jul 26 07:26:23 2019 ---> 1.9 uA
*Fri Jul 26 16:27:50 2019 ---> 1.9 uA
*Sat Jul 27 01:29:52 2019 ---> 1.9 uA
Attached is also the output of the HV monitor (~/HV_7.root) where the leakage current increase is visible. I just did not manage to draw it as a function of absolute time because the date stored in the file seems wrong. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|