Message ID: 39     Entry time: Tue Mar 5 16:20:11 2019
Author: iulian Marlene 
ExpNbr: TEST 
Type: MUST2 
Category: Software 
Status: Fixed 
Target-Source: N/A 
Subject: run 77 Gamme 600 ns Time Calibration run (Calimero) on MUVI 3 MM1 X . good run 
Run 76


Calimero on Muvi3 MM1(X)

Gamme 600 ns

Procedure:

0) HT on the detectors Off
1) Apply Das config
2) Put Das offline
3) Start Geco
4) Activate on each of the sector to calibrate the X side (see bellow image)
5) execute Calimero.exe in calimero directory
6) modify the calimero config on Geco (see bellow near config button)
7) Start the run in RCC
8) Start calimero (start button in Geco)
Attachment 1: run77_calimero_mm1X_spectra.png  94 kB  | Show | Show all