MUST2 SAMURAI_2018 S3-LEB-LPC SUPERNEMO MUGAST EXPAND SCALP GALATRON HiCARI VELO
  MUGAST  Not logged in MUGAST
Message ID: 18     Entry time: Thu Feb 21 17:32:42 2019
Author: iulian, freddy 
ExpNbr: TEST 
Type: GRIT 
Category: Hardware 
Status: Fixed 
Target-Source: 3-alpha 
Subject: run 33 3 alfa with MG1,MG3,MG4,MG5,MG7,MG9 short run 
First test of the repaired 2xcables. The repaired cables were placed between MUVI1_2 <=> MG4 & MUVI1_4 <=>MG11 (longer cable)
Correspondance for the rest is the same as for run 30. The mapping is the same for all detectors (including anulaire detector).

Alpha source run for trapezoid detectors + anulaire detector with pedestals aligned and zero suppression. Cooling at 15°C.
Thresholds at 500 keV.

On MUVI1:

1--> MG5 (150V)
2--> MG4 (95V)
3--> MG3 (160V)
4--> MG11 (60V)

On MUVI2:
5--> MG7 (120V)
6--> MG1 (140V)
7--> --
8--> --

TRIGGERS are:
GMT ch1 --> OR MUVI1
GMT ch2 --> OR MUVI2
GMT ch3 --> OR MUVI3 (nothing on it)
Attachment 1: 3alpha_sourceMG1MG4MG7MG9XY_run33.png  106 kB  | Hide | Hide all
3alpha_sourceMG1MG4MG7MG9XY_run33.png
Attachment 2: 3alpha_sourceMG3MG5XY_run33.png  88 kB  | Hide | Hide all
3alpha_sourceMG3MG5XY_run33.png

ELOG Home Page